Gw-instek AFG-3000 Series User Manual Page 206

  • Download
  • Add to my manuals
  • Print
  • Page
    / 304
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 205
AFG-3000 Series User Manual
206
The offset is also determined by the output
termination (50 or high impedance). If the offset
has been set and the output termination has
changed from 50 to high impedance, the offset
will double. Changing the output termination from
high impedance to 50 will half the offset.
SOURce[1]:APPLy:SINusoid
Source Specific
Command
Description
Outputs a sine wave from the selected channel
when the command has executed. Frequency,
amplitude and offset can also be set.
Syntax
SOURce[1]:APPLy:SINusoid [<frequency>
[,<amplitude> [,<offset>] ]]
Parameter <frequency>
1μHz~80MHz(3081)/
50MHz(3051)
<
amplitude
>
10mV~10V (50) (3.536
Vrms)
<offset>
0~4.99V (50)
Example SOUR1:APPL:SIN 2KHZ,MAX,MAX
Sets frequency to 2kHz and sets the amplitude and
offset to the maximum.
SOURce[1]:APPLy:SQUare
Source Specific
Command
Description
Outputs a square wave from the selected channel
when the command has executed. Frequency,
amplitude and offset can also be set. The duty
cycle is set to 50%.
Syntax
SOURce[1]:APPLy:SQUare [<frequency> [,<amplitude>
[,<offset>] ]]
Parameter <frequency>
1μHz~80MHz(3081)/
50MHz(3051)
<
amplitude
>
10mV~10V (50)
Page view 205
1 ... 205 206 207 ... 304

Comments to this Manuals

No comments