REMOTE CONTROL
97
Function
Queries the current measurement range
Example
:SENSe:CURRent:RANGe?
Returns the current measurement range.
Command
:SENSe[1]:CURRent[:DC]:RANGe:AUTO <b>
Function
Turns the automatic range function.
Description
<b>
0/OFFTurn off.
1/ONTurn on.
Example
:SENSe:CURRent:RANGe:AUTO ON
Turns the automatic range function on.
Command
:SENSe[1]:CURRent[:DC]:RANGe:AUTO?
Function
Queries the state of the automatic range function.
Example
:SENSe:CURRent:RANGe:AUTO?
Returns the status of the automatic range function.
Command
:SENSe[1]:PCURrent:AVERage <NRf>
Function
Sets the averaging number for pulse current
measurements.
Description
NRf
1-100 or 1-5000(pulse current
digitization)
Example
:SENSe:PCURrent:AVERage 5
Sets the average number to 5.