Hand-held LCR Meter LCR-914/915/916 Series USER MANUAL GW INSTEK PART NO. 82CR-91600MA1 ISO-9001 CERTIFIED MANUFACTURER
LCR-914/915/916 User Manual 8 Making 4-Wire Measurements with the 5-Wire Terminal Making 2-Wire Measurements with the Alligator Clip Set
LCR-914/915/916 User Manual 9 Measuring L/C/R/DCR • Press the L/C/R/DCR button to select the measurement function. • Press the L/C/R/DCR button
LCR-914/915/916 User Manual 10 Measuring D/Q/ESR/θ • Press the D/Q/ESR/θbutton to select the measurement function.
LCR-914/915/916 User Manual 11 Selecting the Test Frequency • Press the FREQ button to select the test frequency. • The selectable frequency r
LCR-914/915/916 User Manual 12 Selecting the Series / Parallel Measurement Function • At the L/C/R measuring function, it defaults to Auto Se
LCR-914/915/916 User Manual 13 Selecting the Display Count • Press the 2000 /20000 button to select the display count. Zero The Zero
LCR-914/915/916 User Manual 14 Display Hold • Press the HOLD button to hold the reading on the meter, press the button again to return. D
LCR-914/915/916 User Manual 15 Calibratation In order to achieve the best measurement results, calibration is essential. To calibrate the meter,
LCR-914/915/916 User Manual 16 When “OPEn” appears on the sub display, make the terminal or the SMD test probe open, and press t
LCR-914/915/916 User Manual 17 When “Srt” appears on the sub display, short the terminals or the SMD test probe, and then press the CAL button to
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied,
LCR-914/915/916 User Manual 18 Sorting To check the accuracy of a part, press the SORTING button to enter the sorting mode. The sorting result a
LCR-914/915/916 User Manual 19 The default sorting standard value is the current value, and the default tolerance is ±1.0%. Setup Sorting
LCR-914/915/916 User Manual 20 Battery Replacement Refer to the following figure to replace the batteries : External Power Sour
LCR-914/915/916 User Manual 21 Specifications General Specifications Maximum voltage applied to any terminal: 30VDC or 30VAC rms Display: 2000/
LCR-914/915/916 User Manual 22 Shock Vibration: Sinusoidal vibration per MIL-T- 28800E (5 ~ 55 Hz, 3g maximum). Drop Protection: 4 feet drop to h
LCR-914/915/916 User Manual 23 (4) Inductance Frequency Range Accuracy 100Hz 120Hz 20.000mH ± (0.5% + 5d) [2] 200.00mH ± (0.2% + 2d) 2000.0mH 20.
LCR-914/915/916 User Manual 24 (5) Capacitance Frequency Range Accuracy 100Hz 120Hz 2000.0pF ± (0.5% + 5d) [2] 20.000nF ± (0.2% + 2d) 200.00nF 2000
LCR-914/915/916 User Manual 25 (6) Resistance Frequency Range Accuracy 100Hz 120Hz 200.00Ω ± (0.2% + 5d) 2.0000kΩ ± (0.2% + 2d) 20.000kΩ 200.00kΩ 2
LCR-914/915/916 User Manual 26 (7) DCR Range Resolution Accuracy 200.00Ω 0.01Ω ± (0.2% + 5d) 2.0000kΩ 0.0001kΩ ± (0.2% + 2d) 20.000kΩ 0.001kΩ 200.0
LCR-914/915/916 User Manual 1 Read First Safety Information Understand and follow the operating instructions carefully. Use the meter only
LCR-914/915/916 User Manual 2 CAUTION ˙ Never connect a source of voltage that could result in damage to the meter and the equipment under tes
LCR-914/915/916 User Manual 3 Meter Description Front Panel Illustration 1. LCD display: 20000/2000 counts. 2. Function buttons. 3. 5
LCR-914/915/916 User Manual 4 Package Contents 1. 5V AC Adapter (only LCR-916). 2. USB Cable (only LCR-916). 3. Shorting Cube. 4. SMD Test Probe
LCR-914/915/916 User Manual 5 Measuring Principles tanD1QRXXtanC1fC21XLfL2XRXXtan)X(XRZ)Xj(XRECL1CLCL1CL2CL
LCR-914/915/916 User Manual 6 Phase Drawing
LCR-914/915/916 User Manual 7 Making 5-Wire Measurements with the SMD Test Probe