LEAKAGE CURRENT TESTER
GLC-9000
QUICK START GUIDE
GWINSTEK PART NO.82LC-90000MB1
ISO-9001 CERTIFIED MANUFACTURER
Page view 0
1 2 3 4 5 6 ... 30 31

Summary of Contents

Page 1 - LEAKAGE CURRENT TESTER

LEAKAGE CURRENT TESTER GLC-9000 QUICK START GUIDE GWINSTEK PART NO.:82LC-90000MB1 ISO-9001 CERTIFIED MANUFACTURER

Page 2

GLC-9000 Quick Start Guide 10 5. Patient Leakage Current III Refers to the current that flows through a human body that is directly connected to

Page 3 - Table of Contents

INTRODUCTION 11 Measurement Modes There are a number of leakage current measurement tests covering general electrical equipment and medical electr

Page 4

GLC-9000 Quick Start Guide 12 Operation  Auto/Manual/Programmable Single fault conditions and power supply polar

Page 5 - INTRODUCTION

INTRODUCTION 13 Remote Interface There are a variety of remote control interfaces including: RS-232, USB (Host/Device), GPIB and the EXT I/O conne

Page 6 - Front Panel

GLC-9000 Quick Start Guide 14 Power and Probe Connection Mains Power 1. Ensure the power is switched off from the front panel. 2. Insert the

Page 7

INTRODUCTION 15 Caution If network B (MD B) is selected an isolation transformer that outputs 110% of the rated voltage specified for the EUT is r

Page 8 - Rear Panel

GLC-9000 Quick Start Guide 16 The arrow above shows the contact point for the test lead. Alligator Clips 1. Plug a test lead into the rear pane

Page 9 - Leakage Current Modes

INTRODUCTION 17 Preparation Voltage Line Selection Before power is turned on, ensure the line voltage is correct for the environment. The line

Page 10 - Key Features

GLC-9000 Quick Start Guide 18 Power Up Press the power switch to turn on the power. The system will enter the measurement interface after a quick

Page 11

INTRODUCTION 19 Interface Max : Min: Name: 10.10 mAClass: I Network: E Save RecallHard Copy NetworkClass LeakageLimit SystemAC+DCCurrentAUTOM

Page 12

This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied,

Page 13

GLC-9000 Quick Start Guide 20 Judge: Shows the test result judgment. Set Limits Upper Limit: Displays the upper test limit. Lower Limit: Display

Page 14 - Power and Probe Connection

INTRODUCTION 21 Indicates that the touch panel is currently unlocked. Press the Lock key to lock the front panel. Indicates the front panel is

Page 15

GLC-9000 Quick Start Guide 22 Save a screen image (BMP) Save settings. Manual measurements can be saved in real-time. Recall settings

Page 16

OPERATION 23 OPERATION Measurement Terminals When a measurement network is selected, different measuring terminals are required for each test an

Page 17 - Preparation

GLC-9000 Quick Start Guide 24 Medical Equipment MD-F CLASS I CLASS II Internal Power Type B Type F* Type B Type F* Type B Type F* Earth le

Page 18

OPERATION 25 Leakage Current Testing Connections Background Confirm all settings including MD, leakage current mode, measurement time, upper and

Page 19 - Interface

GLC-9000 Quick Start Guide 26 The T3 terminal is high voltage. Avoid contact with the terminal. The T3 terminal should not be connected with an e

Page 20

OPERATION 27 Enclosure and Line Leakage Current Non-medical type (General electrical equipment) The T2 terminal is high voltage. Avoid contact

Page 21

GLC-9000 Quick Start Guide 28 Medical network (Internal power supply) Patient Leakage Current II Type B Medical Network The T3 terminal is h

Page 22

OTHER 29 OTHER EUT Voltage and Current Check Background The EUT voltage and current check tests voltage, current and power consumption. Panel Ope

Page 23 - OPERATION

Table of Contents 3 Table of Contents INTRODUCTION ... 5Introduction ...

Page 24 - Medical Equipment

GLC-9000 Quick Start Guide 30 Voltage Measurement Background From the System menu, the meter function can measure different types of voltages: AC

Page 25

OTHER 31 1. To set the safety extra low voltage, press the SELV key. 2. To set the SELV use the number pad and press Confirm. SELV Range

Page 27

INTRODUCTION 5 INTRODUCTION This quick start guide is designed to help those who are not familiar with the GLC-9000 to quickly learn the fundament

Page 28

GLC-9000 Quick Start Guide 6 Front Panel Item Description 1. POWER The power switch turns the power on or off. : ON

Page 29

INTRODUCTION 7 5. Warning Indicator The warning indicator lights up when high voltages are produced from terminals T1, T2 or T3. The warning ind

Page 30 - Voltage Measurement

GLC-9000 Quick Start Guide 8 Rear Panel ITEM Description 1. EUT AC IN EUT AC inlet. AC voltage range: 85V~ 250V AC, 50~60Hz 10A Max 2. R

Page 31 - System Settings

INTRODUCTION 9 Leakage Current Modes 1. Earth Leakage Current Refers to the current that flows through a protective grounding wire to earth.