Measurement Commands
Command
:FETCh?
Function
Returns the last readback value.
Response Time
Maximum: 16ms.
Example
:FETCh?
Returns the last readback value.
Command
:FETCh:ARRay?
Function
Returns the last array readback values.
Response Time
Maxium: 16ms
Example
:FETCh:ARRay?
Returns the last array readback values
Command
:READ?
Function
Triggers a read operation and returns the read
values.
Response time
Maximum: 32ms
Example
:READ?
Returns the read values.
Command
:READ:ARRay?
Function
Triggers a new array. Returns the read array
values.
Response time
Max: 32ms
Example
:READ:ARRay?
Triggers a new array. Returns the read array
values.