Gw-instek PPH-1503 User Manual Page 51

  • Download
  • Add to my manuals
  • Print
  • Page
    / 137
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 50
BASIC OPERATION
51
Timeout
Pulse timeout
When long integration measurement is
selected and the unit doesn’t detect a
pulse after a certain amount of time
(pulse timeout time), the “No Pulse”
message will be displayed on the LCD.
This function is only applicable if rising
or falling edge is selected as the edge
trigger; the Trig On Neither trigger
setting has no pulse timeout. The pulse
timeout has a range of 1~63 seconds.
TrigLeve[4]
Trigger level.
When the rising or falling edge trigger is
selected for long integration current
measurement, a pulse must first be
detected. The trigger level refers to
minimum pulse level required for a
pulse to be detected. For example if the
trigger level is set to 2A, pulses that are
≤2A will be detected. Pulses <2A will be
ignored. The trigger level range is 0~5A.
This setting only applies to long current
integration measurements.
RecallSetup
Recalls pre-saved setups. A total of 6
setups can be recalled: RST/SAV0 ~
SAV4. See page 60 for details.
Page view 50
1 2 ... 46 47 48 49 50 51 52 53 54 55 56 ... 136 137

Comments to this Manuals

No comments