Gw-instek PSU-Series Programming User Manual User Manual

Browse online or download User Manual for Equipment Gw-instek PSU-Series Programming User Manual. GW Instek PSU-Series Programming User Manual User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 128
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews

Summary of Contents

Page 1 - Programmable DC Power Supply

Programmable DC Power Supply PSU Series PROGRAMMING MANUAL Revision 1.2 July 2014 ISO-9001 CERTIFIED MANUFACTURER

Page 2

PSU Programming Manual 10 Main Features Performance  High power density: 1500W in 1U  Universal input voltage 85~265Vac, continuous operation

Page 3 - Table of Contents

PSU Programming Manual 100 Query Example SYST:VERS? >1999.9 :SYSTem:REBoot Set Description Reboots the PSU system. Syntax :SYSTem:REBoot IEEE

Page 4 - SAFETY INSTRUCTIONS

REMOTE CONTROL 101 Syntax Query Syntax *ESE <NR1> *ESE? Parameter <NR1> 0~255 Return parameter <NR1> Returns the bit sum of the

Page 5 - Safety Guidelines

PSU Programming Manual 102 Return parameter 1 Returns 1 when all the outstanding commands have completed. *RCL Set Description Recalls the conten

Page 6

REMOTE CONTROL 103 Parameter <NR1> 0~255 Return parameter <NR1> Returns the bit sum of the Service Request Enable register. *STB Quer

Page 7

PSU Programming Manual 104 Status Register Overview To program the PSU power supply effectively, the Status registers need to be understood. This

Page 8 - GETTING STARTED

REMOTE CONTROL 105 The Status Registers MSSConditionOV (Over-Voltage) 0PTR/NTP Event EnableOC (Over-Current) 1Not Used 2POW (AC Power Off) 3OT

Page 9 - PSU Series Overview

PSU Programming Manual 106 Questionable Status Register Group Overview The Questionable Status Register Group indicates if any protection modes o

Page 10 - Main Features

REMOTE CONTROL 107 OT (Over Temperature) Over temperature protection has been tripped 4 16 VL (Voltage Limit) Voltage limit has been reached 8 2

Page 11 - Accessories

PSU Programming Manual 108 Operation Status Register Group Overview The Operation Status Register Group indicates the operating status of the pow

Page 12

REMOTE CONTROL 109 CV (Constant voltage mode) Indicates if the PSU is in CV mode. 8 256 CC (Constant current mode) Indicates if the PSU is in CC

Page 13 - Appearance

GETTING STARTED 11 Accessories Before using the PSU power supply unit, check the package contents to make sure all the standard accessories are i

Page 14 - Lock/Local

PSU Programming Manual 110 Event Register The PTR/NTR Register will dictate the type of transition conditions will set the corresponding bits in t

Page 15

REMOTE CONTROL 111 Standard Event Status Register Group Overview The Standard Event Status Register Group indicates if any errors have occurred.

Page 16 - Display Area

PSU Programming Manual 112 EXE (Execution Error) The EXE bit indicates an execution error due to one of the following: illegal command parameter,

Page 17

REMOTE CONTROL 113 Status Byte Register & Service Request Enable Register Overview The Status Byte register consolidates the status events of

Page 18 - 234 15678

PSU Programming Manual 114 (ESB) Event Summary Bit. The ESB is the summary bit for the Standard Event Status Register group. 5 32 MSS Bit The MS

Page 19 - REMOTE SENSE

REMOTE CONTROL 115 Error List Command Errors ... 115 Execution Errors ...

Page 20

PSU Programming Manual 116 Error Code Description -100 Command Error This is the generic syntax error for devices that cannot detect more specific

Page 21 - Configuration Settings

REMOTE CONTROL 117 -112 Program mnemonic too long The header contains more that twelve characters (see IEEE 488.2, 7.6.1.4.1). -113 Undefined head

Page 22 - VOLTAGE CURRENT

PSU Programming Manual 118 -141 Invalid character data Either the character data element contains an invalid character or the particular element r

Page 23

REMOTE CONTROL 119 Execution Errors Overview An <error/event number> in the range [ -299 , -200 ] indicates that an error has been detecte

Page 24

PSU Programming Manual 12 Optional Accessories Part number Description PSU-01C Cable for 2 units of PSU-Series in parallel mode connection PSU-0

Page 25 - Configuration Table

PSU Programming Manual 120 -201 Invalid while in local Indicates that a command is not executable while the device is in local due to a hard local

Page 26

REMOTE CONTROL 121 -222 Data out of range Indicates that a legal program data element was parsed but could not be executed because the interpreted

Page 27

PSU Programming Manual 122 or query errors; see the other error definitions in this section. Error Code Description -310 System error Indicates th

Page 28

REMOTE CONTROL 123 Error Code Description -400 Query error This is the generic query error for devices that cannot detect more specific errors. Th

Page 29

PSU Programming Manual 124 APPENDIX PSU Default Settings The following default settings are the factory configuration settings for the power suppl

Page 30 - REMOTE CONTROL

APPENDIX 125 Lock Mode F-19 0:Lock Panel, Allow Output OFF Setup Rear USB Speed F-22 2 = Auto detect GPIB address F-23 8 GPIB Enable/Disable F-24

Page 31 - RMT indicator

PSU Programming Manual 126 Error Messages & Messages The following error messages or messages may appear on the PSU screen during operation.

Page 32 - Function Check

APPENDIX 127 LED ASCII Table Character Set Use the following table to read the LCD display messages. 0 1 2 3 4 5 6 7 8 9 A B C D E

Page 33 - GPIB Remote Interface

PSU Programming Manual 128 INDEX Accessories ... 11 Caution symbol... 4 Cleaning the instru

Page 34

GETTING STARTED 13 Appearance PSU Series Front Panel Lock/Local PROT Function Test Set OutputUnlock ALM_CLR M 1 M 2 M 3Shift: Long PushVSR LAN RM

Page 35

PSU Programming Manual 14 6. Lock/Local Button Lock/LocalUnlock Used to lock all front panel buttons other than the Output Button or it switches t

Page 36

GETTING STARTED 15 11. Shift Button Shift Used to enable the functions that are written in blue characters below the button. 12. Output Button Out

Page 37

PSU Programming Manual 16 PSU Series Display and Operation Panel VSR LAN RMT ERR DLY ALM ISR M 1 M 2 M 3 RUNC CAC VVVOLTAGE CURRENTDisplay Area14

Page 38 - UART Remote Interface

GETTING STARTED 17 24. ISR LED The current slew rate enable. 25. M1 LED Lights in green when the memory value are being recalled or saved. 26. M2

Page 39 - RS 485 / 232

PSU Programming Manual 18 Rear Panel L NAC INPUTDC OUTPUTLANRS 485 / 232OUT INANALOG PROGRAMMINGS LS SLSNCREMOTE SENSE234 156789 10876543210 -

Page 40

GETTING STARTED 19 5. Remote-IN LANRS 485 / 232 Two different types of cables can be used for RS232 or RS485-based remote control. PSU-232: RS232

Page 41 - UART Function Check

This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied,

Page 42

PSU Programming Manual 20 L NAC INPUTDC OUTPUTLANRS 485 / 232OUT INANALOG PROGRAMMINGS LS SLSNCREMOTE SENSE

Page 43

GETTING STARTED 21 Configuration Settings Setting Normal Function Settings The normal function settings (F-01~F-61, F-70~F-76) and F-88 ~ F-89 c

Page 44 - Multi-Unit Function Check

PSU Programming Manual 22 4. Use the current knob to set the parameter for the chosen F setting. Current Press the Voltage knob to save the con

Page 45

GETTING STARTED 23 Setting Power On Configuration Settings Background The Power On configuration settings can only be changed during power up to

Page 46 - Web Server Configuration

PSU Programming Manual 24 Press the Voltage knob to save the configuration setting. ConF will be displayed when it is configuring. Voltage VOLT

Page 47 - LAN indicator

GETTING STARTED 25 Configuration Table Please use the configuration settings listed below when applying the configuration settings. Normal Functi

Page 48 -  Operating area diagram

PSU Programming Manual 26 Internal resistance setting F-08 0~0.03Ω (PSU 6-200) 0~0.104Ω (PSU 12.5-120) 0~0.263Ω (PSU 20-76) 0~1.053Ω (PSU 40-38) 0

Page 49 - Sockets Server Configuration

GETTING STARTED 27 Show MAC Address-2 F-31 0x00~0xFF Show MAC Address-3 F-32 0x00~0xFF Show MAC Address-4 F-33 0x00~0xFF Show MAC Address-5 F-34 0

Page 50 - Socket Server Function Check

PSU Programming Manual 28 UART Address (For TDK) F-76 00 ~ 31 System Settings Factory Set Value F-88 0 = None 1 = Return to factory default settin

Page 51

GETTING STARTED 29 Control Range F-97 0 = 5V [5kΩ], 1 = 10V [10kΩ] External Output Control Function F-98 0 = OFF, 1 = ON Special Function Setting

Page 52

Table of Contents 3 Table of Contents SAFETY INSTRUCTIONS ... 4 GETTING STARTED ...

Page 53

PSU Programming Manual 30 REMOTE CONTROL This chapter describes basic configuration of IEEE488.2 based remote control. Interface Configuration ..

Page 54

REMOTE CONTROL 31 Interface Configuration USB Remote Interface Configuration USB Configuration PC side connector Type A, host PSU side connector

Page 55 - Command Syntax

PSU Programming Manual 32 Function Check Functionality check Invoke a terminal application such as Realterm. To check the COM port No., see the

Page 56

REMOTE CONTROL 33 GPIB Remote Interface Configuration To use GPIB, the optional GPIB option (GW Instek part number: PSU-GPIB) must be installed. T

Page 57 - 1 2 3 4 5

PSU Programming Manual 34 7. The RMT indicator will turn on when a remote connection has been established. RMT indicatorAVVOLTAGE CURRENTVSR LAN

Page 58 - Command List

REMOTE CONTROL 35 Functionality check 1. Start the NI Measurement and Automation Explorer (MAX) program. Using Windows, press: Start>All Prog

Page 59

PSU Programming Manual 36 4. Select the device (GPIB address of PSU) that now appears in the System>Devices and Interfaces > GPIB-USB-HS “

Page 60

REMOTE CONTROL 37 check box is checked, and the terminal character is \n (Value: xA). 11. Click Apply Changes. 10911 12. Click on Input/Output.

Page 61

PSU Programming Manual 38 1213151416 UART Remote Interface Configure UART Overview The PSU uses the IN & OUT ports for UART communication co

Page 62

REMOTE CONTROL 39 PSU-485 RS485 cable with DB9 connector DB-9 Connector Remote IN Port Remarks Pin No. Name Pin No. Name Housing Shield Housing S

Page 63 - :DISPlay:BLINk

PSU Programming Manual 4 SAFETY INSTRUCTIONS This chapter contains important safety instructions that you must follow during operation and storage

Page 64

PSU Programming Manual 40 F-76 = 00~31 UART address if TDK is selected for F-75. 3. The RMT indicator will turn on when a remote connection has

Page 65

REMOTE CONTROL 41 UART Function Check Functionality check Invoke a terminal application such as Realterm. To check the COM port No, see the Devi

Page 66

PSU Programming Manual 42 Multiple Unit Connection Using Local RS485 Bus The PSU power supplies can have up to 31 units daisy-chained together usi

Page 67 - :MEASure[:SCALar]:POWer[:DC]

REMOTE CONTROL 43 4. Terminate the OUT port of the last unit with the end terminal connector included in the PSU-232 or PSU-485 connection kit.

Page 68

PSU Programming Manual 44 F-76 = 00~31 Set the address of each slave to a unique address identifier 7. Multiple units can now be operated at t

Page 69 - :OUTPut:MODE

REMOTE CONTROL 45 Run this query command via the terminal application after the instruments have been configured for multi-unit control. See page

Page 70 - :OUTPut[:STATe][:IMMediate]

PSU Programming Manual 46 Configure Ethernet Connection The Ethernet interface can be configured for a number of different applications. Ethernet

Page 71 - :OUTPut:PROTection:CLEar

REMOTE CONTROL 47 2. Press the Function key to enter the Normal configuration settings. Page 21 Set the following LAN settings: F-36 = 1 Tur

Page 72 - Condition register

PSU Programming Manual 48 F-41 = CCC IP Address part 3 of 4 F-42 = DDD IP Address part 4 of 4 http:// AAA.BBB.CCC.DDD The web browser interf

Page 73 - :STATus:QUEStionable[:EVENt]

REMOTE CONTROL 49 Sockets Server Configuration Configuration This configuration example will configure the PSU socket server. The following conf

Page 74 - :STATus:QUEStionable:ENABle

SAFETY INSTRUCTIONS 5 Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the s

Page 75 - :STATus:PRESet

PSU Programming Manual 50 Socket Server Function Check Background To test the socket server functionality, National Instruments Measurement and A

Page 76

REMOTE CONTROL 51 23 4. Select Manual Entry of Raw Socket from the popup window. 4 5. Enter the IP address and the port number of the PSU.

Page 77 - [:AMPLitude]

PSU Programming Manual 52 6578 9. Next configure the Alias (name) of the PSU connection. In this example the Alias is: PSU_DC1 10. Click finis

Page 78 - [:SOURce]:CURRent:LIMit:AUTO

REMOTE CONTROL 53 1112 13. Click the Configuration icon, 14. Click on I/O Settings. 15. Make sure the Enable Termination Character check box is

Page 79

PSU Programming Manual 54 20. The *IDN? query will return the Manufacturer, model name, serial number and firmware version in the dialog box. GW

Page 80

REMOTE CONTROL 55 Command Syntax Compatible Standard IEEE488.2 Partial compatibility SCPI, 1999 Partial compatibility Command Structure SCPI comm

Page 81 - [:SOURce]:MODE?

PSU Programming Manual 56 Query A query is a simple or compound command followed by a question mark (?). A parameter (data) is returned. Example

Page 82

REMOTE CONTROL 57 Command Forms Commands and queries have two different forms, long and short. The command syntax is written with the short form o

Page 83

PSU Programming Manual 58 <NR1> integers 0, 1, 2, 3 <NR2> decimal numbers 0.1, 3.14, 8.5 <NR3> floating point 4.5e-1, 8.25e+1 &l

Page 84

REMOTE CONTROL 59 Output Commands :OUTPut:DELay:ON ... 68 :OUTPut:EXTernal:LOGic ...

Page 85

PSU Programming Manual 6 Cleaning the PSU  Disconnect the power cord before cleaning.  Use a soft cloth dampened in a solution of mild deterge

Page 86

PSU Programming Manual 60 [:AMPLitude] ... 83 [:SOURce]:VOLTage[0]:LIMit:AU

Page 87

REMOTE CONTROL 61 :SYSTem:PREset ... 99 :SYSTem:VERSion ...

Page 88

PSU Programming Manual 62 Display Commands :DISPlay:MENU[:NAME] ... 62 :DISPlay[:WINDow]:TEXT:CL

Page 89

REMOTE CONTROL 63 :DISPlay[:WINDow]:TEXT[:DATA] SetQuery Description Sets or queries the data text that will be written to the display. Writing t

Page 90 - [:MODE]

PSU Programming Manual 64 Initiate Commands :INITiate:CONTinuous[:TRANsient] ... 64 :INITiate[:IMMediate]:NAME ...

Page 91 - :SYSTem:COMMunicate:ENABle

REMOTE CONTROL 65 :INITiate[:IMMediate][:TRANsient] Set Description This command controls the enabling of output triggers. When a trigger is enab

Page 92

PSU Programming Manual 66 Measure Commands :MEASure[:SCALar]:ALL[:DC] ... 66 :MEASure[:SCALar]:CURRent[:

Page 93 - :SYSTem:COMMunicate:LAN:DHCP

REMOTE CONTROL 67 :MEASure[:SCALar]:POWer[:DC] Query Description Takes a measurement and returns the average output power. Syntax :MEASure[:SCALa

Page 94 - :SELect]

PSU Programming Manual 68 Output Commands :OUTPut:DELay:ON ... 68 :OUTPut:EXTernal:LOGic

Page 95 - :TRANsmit:BITS

REMOTE CONTROL 69 :OUTPut:EXTernal[:STATe] SetQuery Description Sets whether the output will be turned on or off externally. By default this sett

Page 96 - RANsmit:SBITs

SAFETY INSTRUCTIONS 7 Power cord for the United Kingdom When using the power supply in the United Kingdom, make sure the power cord meets the foll

Page 97

PSU Programming Manual 70 :OUTPut:PON[:STATe] SetQuery Description Sets the output state at power-on. This is the equivalent to the F-92 (Output

Page 98 - :SYSTem:ERRor:ENABle

REMOTE CONTROL 71 :OUTPut[:STATe]:TRIGgered SetQuery Description Turns the output on or off when a software trigger is generated. Syntax Query Sy

Page 99 - :SYSTem:VERSion

PSU Programming Manual 72 Status Commands For an overview of all the status registers, their associated register contents and the system diagram,

Page 100

REMOTE CONTROL 73 :STATus:OPERation:ENABle SetQuery Description Sets or queries the bit sum of the Operation Status Enable register. Syntax Quer

Page 101

PSU Programming Manual 74 :STATus:QUEStionable:CONDition Query Description Queries the status (bit sum) of the Questionable Status register. This

Page 102

REMOTE CONTROL 75 Parameter <NR1> 0~32767 Return parameter <NR1> 0~32767 :STATus:PRESet Set Description This command resets the ENA

Page 103

PSU Programming Manual 76 Source Commands [:SOURce]:CURRent:EXTernal:RANGe ... 76 [:SOURce]:CURRent:EXTernal:SOURce ...

Page 104 - Status Register Overview

REMOTE CONTROL 77 Example CURR:EXT:RANG? LOW Returns LOW range. [:SOURce]:CURRent:EXTernal:SOURce SetQuery Description Sets whether constant cur

Page 105 - The Status Registers

PSU Programming Manual 78 [:SOURce]:CURRent[:LEVel]:TRIGgered[:AMPLitude] SetQuery Description Sets or queries the current level in amps when a s

Page 106

REMOTE CONTROL 79 Query Syntax [:SOURce]:CURRent:PROTection:DELay? Parameter <NR2> 0.1~2.0 seconds, where 0=no delay MAX The maximum allowe

Page 107

PSU Programming Manual 8 GETTING STARTED This chapter describes the power supply in a nutshell, including its main features and front / rear panel

Page 108

PSU Programming Manual 80 Parameter OFF | 0 Turns the OCP off. ON | 1 Turns the OCP on. Return parameter <bool> Returns the over current pr

Page 109

REMOTE CONTROL 81 [:SOURce]:CURRent:SLEW:FALLing SetQuery Description Sets or queries the falling current slew rate. This is only applicable for

Page 110

PSU Programming Manual 82 Syntax [:SOURce]:RESistance[:LEVel][:IMMediate][:AMPLitude] {<NR2>(OHM)|MINimum|MAXimum} Query Syntax [:SOURce]:RE

Page 111

REMOTE CONTROL 83 Example VOLT:EXT:SOUR? NONE External control is not used. [:SOURce]:VOLTage[0][:LEVel][:IMMediate][:AMPLitude] SetQuery Descr

Page 112

PSU Programming Manual 84 [:SOURce]:VOLTage[0]:LIMit:AUTO SetQuery Description Sets whether to limit the voltage setting so that it does not exce

Page 113 - Error Que

REMOTE CONTROL 85 [:SOURce]:VOLTage[0]:PROTection[:LEVel] SetQuery Description Sets or queries the overvoltage protection level. Syntax [:SOURce]

Page 114

PSU Programming Manual 86 [:SOURce]:VOLTage[0]:SLEW:RISing SetQuery Description Sets or queries the rising voltage slew rate. This is only applic

Page 115 - Error List

REMOTE CONTROL 87 System Function Command :SYSTem:CONFigure:BEEPer[:STATe] ... 88 :SYSTem:CONFigure:BLEeder[:STATe]...

Page 116

PSU Programming Manual 88 :SYSTem:CONFigure:BEEPer[:STATe] SetQuery Description Sets or queries the buzzer state on/off. Syntax Query Syntax :SY

Page 117

REMOTE CONTROL 89 3 External resistance control; 10kΩ = Io min, 0kΩ = Io max. :SYSTem:CONFigure:VOLTage:CONTrol SetQuery Description Sets or que

Page 118

GETTING STARTED 9 PSU Series Overview Series lineup The PSU series consists of 5 models, covering a number of different current, voltage and po

Page 119 - Execution Errors

PSU Programming Manual 90 :SYSTem:CONFigure:PROTection:RECovery SetQuery Description Sets or queries how the OHP, FAN, AC-FAIL, and SD alarms are

Page 120

REMOTE CONTROL 91 Parameter HIGH | 0 Active high LOW | 1 Active low Return Parameter <NR1> Returns the logic setting. :SYSTem:COMMunicate:E

Page 121 - Device Specific Errors

PSU Programming Manual 92 Example SYST:COMM:GPIB:SELF:ADDR 15 Sets the GPIB address to 15. :SYSTem:COMMunicate:LAN:IPADdress SetQuery Description

Page 122 - Query Errors

REMOTE CONTROL 93 :SYSTem:COMMunicate:LAN:SMASk SetQuery Description Sets or queries the LAN subnet mask. Note: the setting will only be valid af

Page 123

PSU Programming Manual 94 :SYSTem:COMMunicate:LAN:DNS SetQuery Description Sets or queries the DNS address. Note: the setting will only be valid

Page 124 - APPENDIX

REMOTE CONTROL 95 :SYSTem:COMMunicate:SERial[:RECeive] :TRANsmit:BAUD SetQuery Description Sets or queries the UART baud rate. Note: the setting

Page 125

PSU Programming Manual 96 :SYSTem:COMMunicate:SERial[:RECeive]:TRANsmit:PARity SetQuery Description Sets or queries the parity of the UART connec

Page 126

REMOTE CONTROL 97 :SYSTem:COMMunicate:USB:FRONt:STATe Query Description Queries the front panel USB-A port state. Query Syntax :SYSTem:COMMunicat

Page 127

PSU Programming Manual 98 Return parameter 0 <NR1>Absent 1 <NR1>Connected to the PC :SYSTem:ERRor Query Description Queries the erro

Page 128 - PSU Programming Manual

REMOTE CONTROL 99 Syntax :SYSTem:ERRor:ENABle :SYSTem:LANGuage:EMULation SetQuery Description Sets or queries the command language. Syntax Quer