PSS-3203/2005 PROGRAMMABLE POWER SUPPLYUSER MANUAL ⎯ ⎯iCONTENTS PAGE1. PRODUCT INTRODUCTION...1
PSS-3203/2005 PROGRAMMABLE POWER SUPPLYUSER MANUAL⎯ 14 ⎯6. MAINTENANCEThe following instructions are executed by qualified personnel only.To avo
PSS-3203/2005 PROGRAMMABLE POWER SUPPLYUSER MANUAL⎯ 16 ⎯[Step 2.1]Now input the voltage value (Min) measured by DMM with the knoband press [ENTE
PSS-3203/2005 PROGRAMMABLE POWER SUPPLYUSER MANUAL⎯ 18 ⎯When above calibration steps has been confirmed correctly, switchthe calibration selecti
PSS-3203/2005 PROGRAMMABLE POWER SUPPLYUSER MANUAL⎯ 20 ⎯7-2.The Operation of the whole circuitEach output channel of PSS-series has a reference
PSS-3203/2005 PROGRAMMABLE POWER SUPPLYUSER MANUAL ⎯ ⎯iiiFOR UNITED KINGDOM ONLYNOTE: This lead/appliance must only be wired by competent per
PSS-3203/2005 PROGRAMMABLE POWER SUPPLYUSER MANUAL ⎯ ⎯vDeclaration of ConformityWeGOOD WILL INSTRUMENT CO., LTD.No. 7-1, Jhongsing Rd., Tuche
PSS-3203/2005 PROGRAMMABLE POWER SUPPLYUSER MANUAL⎯ 2 ⎯4) The function of over voltage/current/temperature protection.5) Intelligent control fan
PSS-3203/2005 PROGRAMMABLE POWER SUPPLYUSER MANUAL⎯ 4 ⎯Interface RS232 (GPIB interface option)Power SourceAC100V, 120V, 220V±10%, 230V +10%/-6%
PSS-3203/2005 PROGRAMMABLE POWER SUPPLYUSER MANUAL⎯ 6 ⎯3-3.EnvironmentThe normal ambient temperature range of this instrument is from 0° to 40°C
PSS-3203/2005 PROGRAMMABLE POWER SUPPLYUSER MANUAL⎯ 8 ⎯Figure 4-2 Rear Panel PSS-3203/2005 PROGRAMMABLE POWER SUPPLYUSER MANUAL⎯ 9 ⎯1. Powe
PSS-3203/2005 PROGRAMMABLE POWER SUPPLYUSER MANUAL⎯ 10 ⎯5. OPERATION METHOD5-1. Output Voltage/Current SettingAt first, ensure that the window n
PSS-3203/2005 PROGRAMMABLE POWER SUPPLYUSER MANUAL⎯ 12 ⎯--Over Current Protection Setting:Set to OCP SET window by pressing [MENU], Turn on/off