GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ ⎯iCONTENTS PAGE1. PRODUCT INTRODUCTION...1-1.Descr
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 14 ⎯(10). ◄ MEMO- 9 ► —SAVE/RECALL(GOS-6051/6031)The instrument contains 10 non-volatile memories
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 16 ⎯(15) ALT/CHOPThe pushbutton has two functions, which are required and available onlywhen both
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 18 ⎯(23)CH1-X—Input BNC socketThis BNC socket is the signal input for channel 1. In X-Y mode, sig
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 20 ⎯(28)×1/MAGPressing the pushbutton the select the sweep time between ×1 (normal)and MAG (magni
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 22 ⎯will be displayed.Use this mode when effecting synchronization to a very low frequencysignal
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 24 ⎯(34)COUPLING—Pressing the pushbutton to select the trigger coupling. The actual setting isind
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 26 ⎯4-2.Rear PanelThe rear panel provides input power and additional signal connections.(38)Line
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 28 ⎯5. OPERATION METHODThis section contains basic operation information and techniques that shou
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 30 ⎯5-2.Connecting Input SignalsGroundingThe most reliable signal measurements are made when the
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 32 ⎯shown in figure 5-2. If either probe needs to be adjusted, proceed the step 8.If either probe
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ ⎯iiiFOR UNITED KINGDOM ONLYNOTE: This lead/appliance must only be wired by competent personsWARN
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 34 ⎯4. Set both CH1 and CH2 COUPLING to GND.5. Use the CH1 and CH2 POSITION controls to align bot
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 36 ⎯Displaying the sum or difference of CH1 and CH2To display the algebraic sum or difference of
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 38 ⎯Magnifying Waveform EventsUse the MAG pushbutton to view small portions of a waveform as whic
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 40 ⎯Operating Hold off time ControlWhen the measured signal is a complex waveform with two or mor
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 42 ⎯The VERT-MODE triggering is not possible when the signal is applied only toone channel as sho
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 44 ⎯5-6.Measurement Application (GOS-6051/6031)The oscilloscope has a cursor measurement system f
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 46 ⎯6.MAINTENENCEThe following instructions are executed by qualified personnel only. To avoidele
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 48 ⎯7.Block Diagram
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ ⎯vDeclaration of ConformityWeGOOD WILL INSTRUMENT CO., LTD.No.7-1, Jhongsing Road, Tucheng City,
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 2 ⎯1-2.FeaturesAdditionally, the oscilloscope offers several other features:1) High intensity and
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 4 ⎯2.TECHNICAL SPECIFICATIONSType6-inch rectangular type with internal graticule;0%, 10%, 90% and
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 6 ⎯Panel Setting DisplayCH1/CH2 sensitivity, sweep time, trigger conditionPanel Setting Save &
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 8 ⎯3-3.EnvironmentThe normal ambient temperature range of this instrument is from 0° to 40°C(32°
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 10 ⎯Front panel of GOS-6031/6051Front panel of GOS-6030/6050 GOS-6051/6050/6031/6030 OSCILLOSCOP
GOS-6051/6050/6031/6030 OSCILLOSCOPEUSER MANUAL⎯ 12 ⎯(3) INTEN—Control knobThe control knob is used for adjusting the traces intensity. Turning t