Gw-instek GDS-1000A Series User Manual User Manual

Browse online or download User Manual for Equipment Gw-instek GDS-1000A Series User Manual. GW Instek GDS-1000A Series User Manual User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 135
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews

Summary of Contents

Page 1 - Digital Storage Oscilloscope

Digital Storage Oscilloscope GDS-1000A Series USER MANUAL GW INSTEK PART NO. 82DS-1102AMD1 ISO-9001 CERTIFIED MANUFACTURER

Page 2

GDS-1000A Series User Manual 10 (Pollution Degree) EN 61010-1:2001 specifies pollution degrees and their requirements as follows. The oscilloscop

Page 3 - Table of Contents

GDS-1000A Series User Manual 100 2. If the file/folder still needs to be deleted, press Delete again to complete the deletion. To cancel the del

Page 4

SAVE/RECALL 101 Quick Save (HardCopy) Background The Hardcopy key works as a shortcut for printing screen images directly to a printer or to save

Page 5 - TABLE OF CONTENTS

GDS-1000A Series User Manual 102 4. Press Function repeatedly to select Save Image or Save All. 5. To invert the color in the display image,

Page 6

SAVE/RECALL 103 Save This section describes how to save data using the Save/Recall menu. File type/source/destination Item Source Destination Pa

Page 7 - SAFETY INSTRUCTIONS

GDS-1000A Series User Manual 104 Saving the panel settings Procedure 1. (For saving to an external SD card) Insert the card into the slot. 2.

Page 8 - Safety Guidelines

SAVE/RECALL 105 File utilities To edit SD card contents (create/ delete/ rename files and folders), press File Utilities. For details, see page98.

Page 9

GDS-1000A Series User Manual 106 SD Normal Save to the SD card with a 4k waveform memory length. SD 1M Save to the SD card with a 1M waveform

Page 10

SAVE/RECALL 107 2. Press the Save/Recall key twice to access the Save menu. Save/Recall Save/Recall 3. Press Save Image. 4. Press Ink Saver

Page 11 - SAFETY INSTRUCTIONS

GDS-1000A Series User Manual 108 Saving all (panel settings, display image, waveform) Procedure 1. (For saving to an external SD card) Insert th

Page 12 - GETTING STARTED

SAVE/RECALL 109 SD 1M Save to the SD card with a 1M waveform memory length. For 2 channel operation only. SD 2M Save to the SD card with a 2M w

Page 13 -  External trigger input

SAFETY INSTRUCTIONS 11 Power cord for the United Kingdom When using the oscilloscope in the United Kingdom, make sure the power cord meets the follo

Page 14 - Panel Overview

GDS-1000A Series User Manual 110 Recall File type/source/destination Item Source Destination Default panel setup  Factory installed setting 

Page 15

SAVE/RECALL 111 Recalling the default panel settings Procedure 1. Press the Save/Recall key. Save/Recall 2. Press Default Setup. The factory in

Page 16 - TIME/DIV

GDS-1000A Series User Manual 112 Recalling a reference waveform to the display Procedure 1. The reference waveform must be stored in advance. Se

Page 17

SAVE/RECALL 113 3. Press Recall Setup. 4. Press Source repeatedly to select the file source, internal or external memory. Use the Variable kno

Page 18 - Rear Panel

GDS-1000A Series User Manual 114 3. Press Recall Waveform. The display shows the available source and destination options. 4. Press Source re

Page 19 - Display

SAVE/RECALL 115 File utilities To edit the SD card contents (create/ delete/ rename files and folders), press File Utilities. For details, see page

Page 20 - Setting up the Oscilloscope

GDS-1000A Series User Manual 116 4. Use the Variable knob to choose a file name (DSXXXX.BMP). VARIABLE SD card The image file must be placed

Page 21

PRINT 117 PRINT The GDS-1000A is able to print screen images directly to a PictBridge compatible printer. The printed images can use the “Ink Saver

Page 22

GDS-1000A Series User Manual 118 2. Insert the other end of the USB cable into the printer USB port. LINE VOLTAGEAC 10 0 240VFUSE RATINGRANG

Page 23 - QUICK REFERENCE

PRINT 119 10. To change the default page size, press Page Size. Default Default printer page setting. 4 X 6 4 X 6 inches A4 Standard A4

Page 24 - CH1/2 key

GDS-1000A Series User Manual 12 GETTING STARTED The Getting started chapter introduces the oscilloscope’s main features*, appearance, and set up p

Page 25 - Cursor key 2/2

GDS-1000A Series User Manual 120 MAINTENANCE Two types of maintenance operations are available: calibrating the vertical resolution, and compensat

Page 26 - Help key

MAINTENANCE 121 7. The Channel1 calibration will complete in less than 5 minutes. Ch1 calibration 1/3 8. When finished, connect the calibration

Page 27 - Horizontal menu key

GDS-1000A Series User Manual 122 4. Press Wavetype repeatedly to select the standard square wave. 5. Press the Autoset key. The compensation

Page 28 - Math key 2/2 (FFT/FFT rms)

FAQ 123 FAQ • The input signal does not appear in the display. • I want to remove some contents from the display. • The waveform does not update (

Page 29 - Run/Stop key

GDS-1000A Series User Manual 124 The waveform does not update (frozen). Press the Run/Stop key to unfreeze the waveform. See page45 for details.

Page 30 - Save/Recall key 2/10

FAQ 125 The accuracy does not match the specifications. Make sure the device is powered on for at least 30 minutes, within +20°C~+30°C. This is ne

Page 31 - Save/Recall key 4/10

GDS-1000A Series User Manual 126 APPENDIX Fuse Replacement Procedure 1. Remove the power cord and remove the fuse socket using a minus driver.

Page 32 - Save/Recall key 6/10

APPENDIX 127 GDS-1000A Series Specifications The specifications apply when the oscilloscope is powered on for at least 30 minutes under +20°C~+30°C

Page 33 - Save Image

GDS-1000A Series User Manual 128 Common specifications Vertical Sensitivity 2mV/div~10V/Div (1-2-5 increments) Accuracy ± (3% x |Readout|+0.

Page 34 - Save/Recall key 10/10

APPENDIX 129 Cursors and Measurement Voltage Vpp, Vamp, Vavg, Vrms, Vhi, Vlo, Vmax, Vmin, Rise Preshoot/ Overshoot, Fall Preshoot/ Overshoot Ti

Page 35 - Trigger Holdoff

GETTING STARTED 13 Interface  SD/SDHC card interface for saving and recalling data  Calibration output  External trigger input  USB slave

Page 36 - Trigger key 4/6

GDS-1000A Series User Manual 130 Probe Specifications GDS-1062A/1102A/1152A Probe Applicable model & probe GDS-1062A GTP-060A-4* GDS-1102A G

Page 37 - Trigger key 6/6

APPENDIX 131 EC Declaration of Conformity We GOOD WILL INSTRUMENT CO., LTD. No.7-1, Jhongsing Rd., Tucheng City, Taipei County 236, Taiwan GOOD WI

Page 38 - Calibration

GDS-1000A Series User Manual 132 INDEX 2M memory length limits ... 95 AC coupling ... 77 Accumulating waveform

Page 39 - Hardcopy – Save All

INDEX 133 safety instructions ... 9 Equivalent time sampling ... 67 Expand Center ... 77 Expan

Page 40 - Probe compensation

GDS-1000A Series User Manual 134 compensation menu tree ... 40 compensation signal overview ... 48 faq...

Page 41 - Default Settings

INDEX 135 expand waveform ... 77 file format ... 94 invert waveform ...

Page 42 - Built-in Help

GDS-1000A Series User Manual 14 Panel Overview Front Panel LCD display TFT color, 320 x 234 resolution, wide angle view LCD display. Function ke

Page 43 - MEASUREMENT

GETTING STARTED 15 Utility key Utility Configures the Hardcopy function (page101), shows the system status (page88), selects the menu language (pa

Page 44 - Channel 1 off Channel 1 on

GDS-1000A Series User Manual 16 Trigger force key FORCE Acquires the input signal once regardless of the trigger condition at the time (page88). H

Page 45 - Run/Stop

GETTING STARTED 17 External trigger input EXT TRIG Accepts an external trigger signal (page81). Power switch Powers the oscilloscope on or off.

Page 46

GDS-1000A Series User Manual 18 Rear Panel Power cord socket Fuse socket Power cord socket accepts the AC mains, 100 ~ 240V, 50/60Hz. The fuse

Page 47

GETTING STARTED 19 Display Waveform markerVertical status Horizontal status Frequency Trigger conditionWaveform position Trigger status Acquisition

Page 48 - GDS-1152A

December 2010 edition This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manua

Page 49

GDS-1000A Series User Manual 20 Setting up the Oscilloscope Background This section describes how to set up the oscilloscope properly including a

Page 50 - Automatic Measurements

GETTING STARTED 21 6. Connect the probe between the Channel1 input terminal and probe compensation signal output (2Vp-p, 1kHz square wave). 7.

Page 51 - Source 2 first rising edge

GDS-1000A Series User Manual 22 Over CompensationNormalUnder Compensation 11. Setting up the oscilloscope is complete. You may continue with the

Page 52 - Automatic measurement gating

QUICK REFERENCE 23 QUICK REFERENCE This chapter lists the oscilloscope menu tree, operation shortcuts, built-in help coverage, and default factory

Page 53

GDS-1000A Series User Manual 24 Select acquisition mode Normal ~ Peak-Detect Select average number Average Turn Delay on/off Delay On CH1/2 ke

Page 54

QUICK REFERENCE 25 Cursor key 1/2 Turn cursor on/off Cursor Move X1 cursor X1→ VAR Move X2 cursor X2→ VAR Move both X1 and X2 cursor X1X2→ VA

Page 55

GDS-1000A Series User Manual 26 Display key Select waveform type Type Waveform accumulate On/Off Accumulate Refresh accumulation Refresh Set di

Page 56 - Cursor Measurements

QUICK REFERENCE 27 Horizontal menu key Switch from Horizontal Menu to Horizontal Position Menu. Horizontal MENU Select main (default) display Ma

Page 57 - Using the vertical cursors

GDS-1000A Series User Manual 28 Math key 1/2 (+/-/x) Math on/off Math Select math operation type (+/–/x/FFT/FFT rms) Operation Set result posit

Page 58

QUICK REFERENCE 29 Measure key Turn on/off measurement Measure Select measurement type Voltage/Time/Delay Select measurement item VAR or Icon(

Page 59 - Math Operations

TABLE OF CONTENTS 3 Table of Contents SAFETY INSTRUCTIONS ... 7Safety Symbols ...

Page 60

GDS-1000A Series User Manual 30 Save/Recall key 1/10 RecallSetupRecallWaveformDisplayRefs.SaveSetupSaveWaveformSaveImageSaveAllTo RecallSetupTo R

Page 61 - Using the FFT function

QUICK REFERENCE 31 Save/Recall key 3/10 Select other menu Recall Waveform Select waveform source Source→VAR Select waveform destination Destin

Page 62

GDS-1000A Series User Manual 32 Save/Recall key 5/10 Select other menu Display Refs. Turn ref. waveform A on/off Ref.A Turn ref. waveform B on/

Page 63 - CONFIGURATION

QUICK REFERENCE 33 Save/Recall key 7/10 Select other menu Save Waveform Select source Source→VAR Select destination Destination →VAR Save wa

Page 64

GDS-1000A Series User Manual 34 Save/Recall key 9/10 Select other menu Save All Turn on/off ink saver Ink Saver Select destination Destination

Page 65 - Selecting Delay mode

QUICK REFERENCE 35 Trigger key 1/6 TypeEdgeSourceCH1ModeAutoSlope /CouplingHoldoff40.0nsSet toMinimumMENU MENUTrigger TypeTrigger Holdoff Select T

Page 66

GDS-1000A Series User Manual 36 Trigger key 3/6 Select edge trigger type Edge Select trigger source Source Go to slope/coupling menu (page37) S

Page 67

QUICK REFERENCE 37 Trigger key 5/6 Select trigger slope type Slope Select trigger coupling mode Coupling Select frequency rejection Rejection T

Page 68

GDS-1000A Series User Manual 38 Utility key 1/6 Go to hardcopy menu Hardcopy Go to probe compensation menu ProbeComp Select language Language S

Page 69 - Selecting the display grid

QUICK REFERENCE 39 Utility key 3/6 FunctionSave AllPrevious MenuOn/ OffInk SaverOffHardcopy – Save AllMem LengSD NormalSD NormalSD 1M/ SD 2MFast/

Page 70

GDS-1000A Series User Manual 4 Trigger key 2/6 ... 35Trigger key 3/6 ...

Page 71

GDS-1000A Series User Manual 40 Utility key 5/6 Select Hardcopy functionSelect Function Turn on/off Inksaver Ink Saver Go to previous menu Prev

Page 72 - Zoom width

QUICK REFERENCE 41 Default Settings Here are the factory installed panel settings which appear when pressing the Save/Recall key→ Default Setup.

Page 73

GDS-1000A Series User Manual 42 Built-in Help The Help key shows the contents of the built-in help support. When you press a function key, its de

Page 74 - Horizontal Adjustment Menu

MEASUREMENT 43 MEASUREMENT The Measurement chapter describes how to properly observe a signal using the oscilloscope’s basic functions, and how to

Page 75

GDS-1000A Series User Manual 44 Channel iconChannelindicatorChannel 1 off Channel 1 on De-activating a channel To de-activate the channel, press

Page 76 - Vertical View (Channel)

MEASUREMENT 45 Undo optionBefore Autoset After Autoset Undoing the Autoset To undo the Autoset, press Undo (available for a few seconds). Adjustin

Page 77

GDS-1000A Series User Manual 46 Waveform operation Waveforms can be moved or scaled in both the Run and Stop mode. For details, see page70 (Horiz

Page 78

MEASUREMENT 47 Selecting the horizontal scale To select the timebase (scale), turn the TIME/DIV knob; left (slow) or right (fast). TIME/DIV Range

Page 79

GDS-1000A Series User Manual 48 Using the probe compensation signal Background This section introduces how to use the probe compensation signal f

Page 80

MEASUREMENT 49 3. Press ProbeComp. 4. Press Wave type repeatedly to select the wave type. 5. (For only) To change the frequency, press Freq

Page 81 - Trigger

TABLE OF CONTENTS 5 Adjusting the display contrast ... 69Selecting the display grid ...

Page 82

GDS-1000A Series User Manual 50 Automatic Measurements The automatic measurement function measures input signal attributes and updates them in the

Page 83 - Configuring Holdoff

MEASUREMENT 51 Vlo Global low voltage. Vavg Averaged voltage of the first cycle. Vrms RMS (root mean square) voltage. ROVShoot Rise oversho

Page 84 - Configuring the edge trigger

GDS-1000A Series User Manual 52 FRF Time between: Source 1 first rising edge and Source 2 first falling edge FFR Time between: Source 1 first

Page 85

MEASUREMENT 53 3. The measurement results appear on the menu bar, constantly updated. All measurements are derived from the cursor positions. See

Page 86

GDS-1000A Series User Manual 54 Editing a measurement item 3. Press the corresponding menu key (F1~F5) to select the measurement slot to be edi

Page 87

MEASUREMENT 55 7. Press F2 repeatedly to change the channel for Source2. Range CH1, 2, Math View all measurements 8. Press F3 to view all mea

Page 88

GDS-1000A Series User Manual 56 Cursor Measurements Cursor lines, horizontal or vertical, show the precise position of the input waveforms or the

Page 89 - USB Port Interface

MEASUREMENT 57 Moving the horizontal cursors To move the left cursor, press X1 and then use the Variable knob. To move the right cursor, press X2

Page 90 - Remote Control Interface

GDS-1000A Series User Manual 58 Moving the vertical cursors To move the upper cursor, press Y1 and then use the Variable knob. To move the lower

Page 91

MEASUREMENT 59 Math Operations The Math operations can add, subtract, multiply or perform FFT/FFT RMS on the input waveforms. The resulted waveform

Page 92 - System Settings

GDS-1000A Series User Manual 6 Saving the display image ... 106Saving all (panel settings, displa

Page 93

GDS-1000A Series User Manual 60 Suitable for... Single-shot phenomenon (this mode is the same as having no window at all) Blackman FFT window Fr

Page 94 - SAVE/RECALL

MEASUREMENT 61 Using the FFT function Procedure 1. Press the Math key. MATH 2. Press Operation repeatedly to select FFT or FFT RMS. 3. Press

Page 95

GDS-1000A Series User Manual 62 8. To clear the FFT result from the display, press the Math key again. MATH

Page 96

CONFIGURATION 63 CONFIGURATION The Configuration chapter describes how to configure panel settings to make measurements and observations suited to

Page 97 - Setup file format

GDS-1000A Series User Manual 64 Average Multiple data is averaged to form a waveform. This mode is useful for drawing a noise-free waveform. To s

Page 98

CONFIGURATION 65 7. Press Normal. 8. Press Peak-Detect and see that a spike noise is captured. Example The peak detect mode reveals the occas

Page 99

GDS-1000A Series User Manual 66 Delay Off With Delay Off the expansion point and trigger point are always in the same position. Thus when the hor

Page 100

CONFIGURATION 67 Real time vs Equivalent time sampling mode Background The oscilloscope automatically switches between two sampling modes, Real-ti

Page 101 - Save/Recal

GDS-1000A Series User Manual 68 Display The Display section describes how to configure the display settings: drawing type, waveform accumulation,

Page 102

CONFIGURATION 69 Example Accumulation off Accumulation on Adjusting the display contrast Procedure 1. Press the Display key. Display 2. Press

Page 103 - File type/source/destination

SAFETY INSTRUCTIONS 7 SAFETY INSTRUCTIONS This chapter contains important safety instructions that should be followed when operating and storing the

Page 104 - Saving the panel settings

GDS-1000A Series User Manual 70 Horizontal View The Horizontal view section describes how to configure the horizontal scale, position, waveform up

Page 105 - Saving the waveform

CONFIGURATION 71 Selecting the waveform update mode Background The display update mode is switched automatically or manually according to the hori

Page 106 - Saving the display image

GDS-1000A Series User Manual 72 Zooming the waveform horizontally Procedure/ range 1. Press the Horizontal Menu key. MENU 2. Press Window. 3

Page 107

CONFIGURATION 73 Viewing waveforms in the X-Y mode Background The X-Y mode compares the voltage of Channel 1 and Channel 2 waveforms in a single d

Page 108

GDS-1000A Series User Manual 74 Horizontal Adjustment Menu Background The horizontal adjustment menu allows markers to be set at different times

Page 109

CONFIGURATION 75 Navigate markers 7. Press Previous to go to the previous marker. 8. Press Next to go to the next marker.

Page 110

GDS-1000A Series User Manual 76 Vertical View (Channel) The Vertical view section describes how to set the vertical scale, position, bandwidth lim

Page 111

CONFIGURATION 77 AC coupling mode. Only the AC portion of the signal appears on the display. This mode is useful for observing AC waveforms mixed

Page 112

GDS-1000A Series User Manual 78 Procedure 1. Press the Channel key. CH 1 2. Press F5 to toggle between Expand Center and Expand Ground. 3. To

Page 113

CONFIGURATION 79 Limiting the waveform bandwidth Background Bandwidth limitation puts the input signal into a 20MHz (−3dB) low-pass filter. This f

Page 114

GDS-1000A Series User Manual 8 Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or cont

Page 115

GDS-1000A Series User Manual 80 2. Press F4 repeatedly to select voltage or current probes. 3. Use the variable knob to edit the voltage or cur

Page 116

CONFIGURATION 81 Trigger The Trigger function configures the conditions by which the oscilloscope captures the incoming signals. Trigger type Edge

Page 117 - Print (Hardcopy)

GDS-1000A Series User Manual 82 The Auto trigger status appears in the upper right corner of the display. Single The oscilloscope acquires the

Page 118 - LINE VOLTAGE

CONFIGURATION 83 line 1~263 for NTSC, 1~313 for PAL/SECAM Pulse condition (pulse trigger) Sets the pulse width (20ns ~ 10s) and the triggering con

Page 119

GDS-1000A Series User Manual 84 Pressing Set to Minimum sets the Holdoff time to the minimum, 40ns. Note: The holdoff function is automatical

Page 120 - MAINTENANCE

CONFIGURATION 85 7. Press Coupling repeatedly to select the trigger coupling, DC or AC. Range DC, AC 8. Press Rejection to select the freque

Page 121 - Ch1 calibration 1/3

GDS-1000A Series User Manual 86 4. Press Standard repeatedly to select the video standard. Range NTSC, PAL, SECAM 5. Press Polarity repeate

Page 122 - Compensation

CONFIGURATION 87 4. Press Mode repeatedly to select the trigger mode, Auto or Normal. To select the Single trigger mode, press the Single key. S

Page 123

GDS-1000A Series User Manual 88 10. Press Noise Rej to turn the noise rejection on or off. Range On, Off 11. Press Previous menu to go back t

Page 124

CONFIGURATION 89 USB Port Interface The USB port can be set to auto detect, however occasionally the USB host type cannot be detected. The USB Port

Page 125

SAFETY INSTRUCTIONS 9 Power Supply WARNING  AC Input voltage: 100 ~ 240V AC, 47 ~ 63Hz  The power supply voltage should not fluctuate more than

Page 126 - APPENDIX

GDS-1000A Series User Manual 90 Remote Control Interface The Remote control interface section describes how to set up the USB interface for PC con

Page 127 - APPENDIX

CONFIGURATION 91 5. Run this query command via the terminal application. *idn? This command should return the manufacturer, model number, serial

Page 128 - Common specifications

GDS-1000A Series User Manual 92 System Settings The system settings show the oscilloscope’s system information and allow changing the language. Vi

Page 129

CONFIGURATION 93 Procedure 1. Press the Utility key. Utility 2. Press Language repeatedly to select the language.

Page 130 - Probe Specifications

GDS-1000A Series User Manual 94 SAVE/RECALL The save function allows saving display images, waveform data, and panel settings into the oscilloscop

Page 131 - EC Declaration of Conformity

SAVE/RECALL 95 Detail Contains the waveform amplitude and time of each point (4k/1M/2M) relative to the trigger point. Fast Only contains the wav

Page 132

GDS-1000A Series User Manual 96 Note: 2M point memory lengths are only available for time bases slower than 10ns/div on a single channel, and 1 M

Page 133 - INDEX

SAVE/RECALL 97 Setup file format Format xxxx.set (proprietary format) A setup file saves or recalls the following settings. Contents Acquire  mo

Page 134

GDS-1000A Series User Manual 98 Using the SD card file utilities Background When an SD card is inserted into the oscilloscope, file utilities (fi

Page 135

SAVE/RECALL 99 SD card indicator When an SD card is inserted into the oscilloscope, an indicator appears at the right bottom corner of the display.

Comments to this Manuals

No comments