Page 1 - Digital Storage Oscilloscope

Digital Storage Oscilloscope GDS-2000 Series USER MANUAL GW INSTEK PART NO. 82DS-22040ME1 ISO-9001 CERTIFIED MANUFACTURER

Page 2

GDS-2000 Series User Manual 10 GETTING STARTED This chapter describes GDS-2000 in a nutshell, including its main features and front / rear panel i

Page 3 - Table of Contents

GDS-2000 Series User Manual 100 4. Press F3 (Window Zoom). The specified range gets zoomed. The ZOOM indicator appears on the bottom left side o

Page 4

CONFIGURATION 101 3. Press the Horizontal menu key. HORIMENU 4. Press F5 (XY). The display shows two waveforms in X-Y format; Channel 1 as X

Page 5 - SAFETY INSTRUCTIONS

GDS-2000 Series User Manual 102 Vertical View (Channel) This section describes how to set the vertical scale, position, and coupling mode. Move wa

Page 6 - Safety Guidelines

CONFIGURATION 103 Select coupling mode Panel operation 1. Press the Channel key. CH1 2. Press F1 (Coupling) repeatedly to select the coupling mo

Page 7

GDS-2000 Series User Manual 104 Invert waveform vertically Panel operation 1. Press the Channel key. CH1 2. Press F2 (Invert) to invert the wav

Page 8

CONFIGURATION 105 Example BW Limit Off BW Limit On Select probe attenuation level Background A signal probe has an attenuation switch to lower

Page 9 - SAFETY INSTRUCTIONS

GDS-2000 Series User Manual 106 Trigger Trigger configures the condition GDS-2000 captures the incoming signal. Trigger type overview Edge (+Dela

Page 10 - GETTING STARTED

CONFIGURATION 107 Trigger parameter overview Trigger source CH1 ~ 4 Channel 1 ~ 4 input signals Line AC mains signal Ext (For 2CH models only)

Page 11 - GDS-2000 Series Overview

GDS-2000 Series User Manual 108 Video standard (video trigger) NTSC National Television System Committee PAL Phase Alternative by Line SECAM SEqu

Page 12 - Main Features

CONFIGURATION 109 Triggers on AC+DC component. Frequency rejection LF Puts a high-pass filter and rejects the frequency below 50kHz. HF Puts a l

Page 13 - Package Contents

GETTING STARTED 11 GDS-2000 Series Overview Series lineup GDS-2000 series consists of 6 models, divided into 2-channel and 4-channel versions. Mod

Page 14 - Appearance

GDS-2000 Series User Manual 110 Use edge trigger Panel operation 1. Press the Trigger menu key. MENU 2. Press F1 repeatedly to select edge tri

Page 15

CONFIGURATION 111 8. Press F3 (Rejection) to select the frequency rejection mode. Rejection OffF 3 Range LF, HF, Off 9. Press F4 (Noise Rej

Page 16

GDS-2000 Series User Manual 112 4. Press F4 (Ext) repeatedly to select the threshold level for the external trigger input. Ext:TTLF 4 Range T

Page 17 - EXT TRIG

CONFIGURATION 113 Use video trigger Panel operation 1. Press the Trigger menu key. MENU 2. Press F1 repeatedly to select video trigger. The vid

Page 18 - Rear Panel

GDS-2000 Series User Manual 114 Use pulse width trigger Panel operation 1. Press the Trigger menu key. MENU 2. Press F1 repeatedly to select p

Page 19

CONFIGURATION 115 7. Press F1 (Slope) repeatedly to select the trigger slope, which also appears at the bottom of the display. Range Rising e

Page 20 - Display

GDS-2000 Series User Manual 116 System Info / Language / Clock This section describes how to set the interface, beeper, language, time/date, and p

Page 21

CONFIGURATION 117  Spanish  Japanese  Russian  German  Dutch  Polish  Italian  French  Portuguese Panel operation 1. P

Page 22 - Tilt stand

GDS-2000 Series User Manual 118 5. Press F4 (Save) to confirm the value. SaveF 4 6. Press F1 (Date) to switch to the Time setting menu. TimeF

Page 23 - Power up

SAVE/RECALL 119 SAVE/RECALL File format / Utility Display image file format ... 120 Waveform file format ...

Page 24 - First Time Use

GDS-2000 Series User Manual 12 Main Features Performance  High sampling rate: up to 1GS/S real-time, 25GS/s equivalent-time  Deep memory: 25k

Page 25

GDS-2000 Series User Manual 120 File Format/Utility Display image file format Format DSxxxx.bmp or Axxxx.bmp (Windows bitmap format) Contents The

Page 26 - QUICK REFERENCE

SAVE/RECALL 121 Contents: waveform data The waveform data can be used for detailed analysis. It consists of horizontal and vertical position of the

Page 27

GDS-2000 Series User Manual 122 Setup file format Format DSxxxx.set or Axxxx.set (proprietary format) The setup file saves or recalls the followi

Page 28 - Horizontal

SAVE/RECALL 123 USB flash drive file utility Background For USB flash drive, file deletion, folder creation, file/folder rename are available from

Page 29

GDS-2000 Series User Manual 124 4. Use the Variable knob to move the cursor. Press F1 (Select) to go into the folder or go back to the previous

Page 30 - Math key (2/2)

SAVE/RECALL 125 3. When editing is completed, press F4 (Save). A new folder or a new folder/file name is created. SaveF 4 4. Press F5 (Previous

Page 31 - Measure key (2/2)

GDS-2000 Series User Manual 126 Quick Save (HardCopy) Background The Hardcopy key works as a shortcut for saving or printing out information. Har

Page 32 - Run/Stop key

SAVE/RECALL 127 Panel operation 1. Connect the drive to the front or rear panel USB port. Note: Only one host connection, front or rear, is allowe

Page 33 - Save/Recall key (2/9)

GDS-2000 Series User Manual 128 Save File type/source/destination Item Source Destination Panel setup (DSxxxx.set)  Front panel settings  I

Page 34 - Save/Recall key (4/9)

SAVE/RECALL 129 Save panel setting Panel operation 1. (For saving to an external USB flash drive) Connect the drive to the front or rear panel US

Page 35 - Save/Recall key (6/9)

GETTING STARTED 13 Package Contents Check the contents before using GDS-2000. Opening the box Contents  Main unit  Probe set GDS-2062:

Page 36 - Save/Recall key (8/9)

GDS-2000 Series User Manual 130 USB External flash drive, no practical limitation on the amount of file. When saved, the setup file is placed in

Page 37 - Trigger key (1/5)

SAVE/RECALL 131 4. Press F2 (Source). Use the Variable knob to select the source signal. SourceF 2 CH1 ~ CH2 (2CH model) Channel 1 ~ 2 si

Page 38 - Trigger key (3/5)

GDS-2000 Series User Manual 132 USB External flash drive, no practical limitation on the amount of file. When saved, the waveform file is placed

Page 39 - Trigger key (5/5)

SAVE/RECALL 133 3. Press F5 (More). MoreF 5 4. Press F1 (Save Image). The display shows the available file destinations. SaveImageF 1 5. P

Page 40 - Utility key (2/9)

GDS-2000 Series User Manual 134 USB External flash drive, no practical limitation on the amount of file. When saved, the image file is placed in

Page 41 - Utility key (4/9)

SAVE/RECALL 135 4. Press F2 (Save All). The display shows the available file destinations. The following files are saved, contained in a folder.

Page 42 - Utility key (6/9)

GDS-2000 Series User Manual 136 Ink Saver On (normal) Ink Saver Off (inverted) 6. Press F3 (Destination). Use the Variable knob to select t

Page 43 - Utility key (8/9)

SAVE/RECALL 137 Recall File type/source/destination Item Source Destination Default panel setup  Factory installed setting  Current front pa

Page 44 - Utility key (9/9)

GDS-2000 Series User Manual 138 Acquisition Mode: Normal Memory length: 500 Channel Scale: 2V/Div CH1: On, CH2/3/4: Off Coupling: DC Invert: O

Page 45 - Default Settings

SAVE/RECALL 139 Recall reference waveform on the display Panel operation 1. The reference waveform must be stored in advance. See page for wavefo

Page 46 - Built-in Help

GDS-2000 Series User Manual 14 Appearance GDS-2064/2104/2204 Front Panel GDS-2062/2102/2202 Front Panel

Page 47 - MEASUREMENT

GDS-2000 Series User Manual 140 Recall panel setting Panel operation 1. (For recalling from an external USB flash drive) Connect the drive to th

Page 48 - Basic Measurement

SAVE/RECALL 141 5. Press F2 (Source) repeatedly to select the file source, internal memory or external USB. Use the Variable knob to change the m

Page 49

GDS-2000 Series User Manual 142 Recall waveform Panel operation 1. (For recalling from an external USB flash drive) Connect the drive to the fro

Page 50 - Run/Stop

SAVE/RECALL 143 5. Press F2 (Source) repeatedly to select the file source, internal memory or external USB. Use the Variable knob to change the m

Page 51 - Horizontal position/scale

GDS-2000 Series User Manual 144 Recall image Panel operation 1. Connect the USB drive to the front or rear panel USB port. Note: Only one host c

Page 52 - Vertical position/scale

SAVE/RECALL 145 6. To show the image on the display, press F3 (Ref Image) ON or F4 (Recall). RecallF 4Ref Image OnF 3 7. The image appears on t

Page 53 - Probe compensation signal

GDS-2000 Series User Manual 146 PRINT OUT Display printout is also available using proprietary PC software, downloadable from GWInstek website. Ov

Page 54 - Frequency

PRINT OUT 147 2 Configure interface Panel operation 1. Press the Utility key. Utility 2. Press F2 (Interface menu). InterfaceMenuF 2 3. Press

Page 55 - Automatic Measurement

GDS-2000 Series User Manual 148 4. To invert the color for the saved or printed display image, press F2 (Ink Saver) and turn On the Ink Saver. I

Page 56

REMOTE CONTROL CONFIG 149 REMOTE CONTROL CONFIG This chapter describes basic configuration of IEEE488.2 based remote control. For command list, re

Page 57 - Individual mode

GETTING STARTED 15 LCD display TFT color, 320 x 234 resolution, wide angle view LCD display. F1 ~ F5 function keys Activates the functions which

Page 58

GDS-2000 Series User Manual 150 Interface Configuration Configure USB interface USB configuration PC side connector Type A, host GDS-2000 side co

Page 59 - Display All mode

REMOTE CONTROL CONFIG 151 Configure RS-232C interface RS-232C configuration Connector DB-9, Male Baud rate 2400, 4800, 9600, 19200, 38400 Parity

Page 60 - Cursor Measurement

GDS-2000 Series User Manual 152 Range None, Odd, Even 9. Connect the RS-232C cable to the rear panel port: DB-9 male connector. For functional

Page 61

REMOTE CONTROL CONFIG 153 Configure GPIB interface (optional) GPIB module installation The optional GPIB module is available as a separate kit. F

Page 62 - Use vertical cursor

GDS-2000 Series User Manual 154 4. The interface icon at display top changes to GPIB. 5. Press F2 (Address). Use the Variable knob to change

Page 63

REMOTE CONTROL CONFIG 155 USB/RS-232C remote control software Termin al application (USB/RS-232C) Invoke the terminal application such as MTTTY

Page 64 - Math Operation

GDS-2000 Series User Manual 156 BATTERY OPERATION The optional battery allows portable operations such as field applications. Battery packs and re

Page 65 - Operation

BATTERY OPERATION 157 3. Insert the battery packs and close the cover. 4. Turn On the power and make sure the battery icon appears at the top le

Page 66 - Channel1

GDS-2000 Series User Manual 158 MAINTENANCE Two types of maintenance operations are available: calibrate vertical resolution, and compensate the p

Page 67

MAINTENANCE 159 8. The calibration for Channel1 starts and ends automatically, in less than 5 minutes. Ch1 calibration 1/3 9. When finished, co

Page 68 - Unit/Div

GDS-2000 Series User Manual 16 Cursor key Configures and runs cursor measurements (page60). Measure key Configures and runs automatic measuremen

Page 69 - Go-NoGo Test

GDS-2000 Series User Manual 160 3. Press F5 (More) twice. MoreF 5 x2 4. Press F1 (ProbeComp Menu). ProbeCompMenuF 1 5. Press F1 (Wavetype) r

Page 70 - Edit: NoGo when

FAQ 161 FAQ • I pressed the Power (On/Standby) key on the front panel but nothing happens. • I connected the signal but it does not appear on the

Page 71 - Edit: Source signal

GDS-2000 Series User Manual 162 I want to remove the (Measurement result / FFT result / Help contents) from the display. To clear automatic measu

Page 72 - Edit: Template (boundary)

FAQ 163 The display image printout is too dark on the background. Use the Inksaver function which reverses the background color. For details, see

Page 73

GDS-2000 Series User Manual 164 APPENDIX Fuse Replacement Step 1. Take off the power cord and remove the fuse socket using a minus driver. 2

Page 74 - Save &

APPENDIX 165 GPIB Module Installation For GPIB interface and remote control details, see page149. GPIB kit contents  GPIB module  Programming

Page 75

GDS-2000 Series User Manual 166 GDS-2000 Specifications The specifications apply when GDS-2000 is powered on for at least 30 minutes under +20°C~+

Page 76 - Run Go-NoGo test

APPENDIX 167 Trigger Sources CH1, CH2, Line, EXT(2ch model only), CH3, CH4(4ch model only) Modes Auto-Level, Auto, Normal, Single, TV, Edge,

Page 77

GDS-2000 Series User Manual 168 Auto Counter Resolution: 6 digits Accuracy: ±2% Signal source: All available trigger source except the Video tr

Page 78 - Program

APPENDIX 169 Probe Specifications Model-specific GTP-060A Applicable to GDS-2062, GDS-2064 Bandwidth DC ~ 60MHz @ Position x 10 Rise time

Page 79 - Edit program

GETTING STARTED 17 Vertical position knob Sets the vertical position of waveforms (page102). Channel menu key Configures the vertical scale and c

Page 80

GDS-2000 Series User Manual 170 Declaration of Conformity We GOOD WILL INSTRUMENT CO., LTD. No. 7-1, Jhongsing Rd, Tucheng City, Taipei County 23

Page 81 - Run program

INDEX 171 INDEX 2-4 channel difference ... 11 AC coupling ... 103 acquisition ...

Page 82

GDS-2000 Series User Manual 172 safety instruction ... 7 equivalent time sampling ... 90 external trigger ...

Page 83 - CONFIGURATION

INDEX 173 compensation menu tree ... 43 compensation signal overview ... 53 faq ... 162 mem

Page 84

GDS-2000 Series User Manual 174 key overview ... 15 menu tree ... 40 shortcut .

Page 85 - Select acquisition mode

GDS-2000 Series User Manual 18 Rear Panel Power switch Power cord socket Fuse socket Power switch turns the main power On () / Off ( ). Power

Page 86

GETTING STARTED 19 GPIB port (optional) Accepts 24 pin male GPIB connector for remote control (page153). Battery packs (optional) Holds 2 packs of

Page 87

December 2010 This manual contains proprietary information, which is protected by copyrights. All rights are reserved. No part of this manual may b

Page 88 - Mem Leng

GDS-2000 Series User Manual 20 Display PowersourceDisplaymodeWaveformInterface Date/Time orMemory BarTriggerstatusAcquisitionmodeChannelstatusTime

Page 89

GETTING STARTED 21 USB RS-232C GPIB (optional) Date/Time Current date and time (page117). Memory bar The ratio and the position of the disp

Page 90

GDS-2000 Series User Manual 22 Set Up Tilt stand Low angle High angle

Page 91

GETTING STARTED 23 Power up Step 1. Connect the power cord to the rear panel socket. (No need when using the battery). 2. Turn On the main pow

Page 92 - Accumulate waveform

GDS-2000 Series User Manual 24 First Time Use Background This section describes how to connect a signal, adjust the scale, and compensate the pro

Page 93 - Contrast

GETTING STARTED 25 4. Select vector waveform Press the Display key, then F1 (Type) twice to select the vector waveform. DisplayTypeVectorsF 1

Page 94 - Turn Off menu

GDS-2000 Series User Manual 26 QUICK REFERENCE This chapter describes GDS-2000 menu tree, shortcuts to major operations, built-in Help access, and

Page 95 - Horizontal View

QUICK REFERENCE 27 Menu Tree / Operation Shortcuts Convention F1 = Press F1 F1 = Press F1 repeatedly F1 ~ F4 = Select one from F1 to F4 and press

Page 96 - Select horizontal scale

GDS-2000 Series User Manual 28 Auto test/Stop key Auto test/Stop → See Program key (page32) CH1 ~ 4 key On/ Off/ /On/ Offx1/ x10/ x100CH

Page 97 - Select waveform update mode

QUICK REFERENCE 29 Display key Vectors/ DotsDisplayTypeDotsF 1On/ OffAccumulate OffF 2RefreshF 3ContrastF 4F 5/ /DISPLAY Select waveform

Page 98 - (memory bar replaced)

TABLE OF CONTENTS 3Table of Contents SAFETY INSTRUCTIONS ... 5GETTING STARTED ...

Page 99 - Zoom waveform horizontally

GDS-2000 Series User Manual 30 Math key (1/2) MATHOperation+F 1CH1+CH2F 2Position0.00 DivF 4Unit/Div2VF 5MATH-12div ~ +12div(4CH)CH1+CH2/CH3+CH4(

Page 100 - Show waveform in X-Y mode

QUICK REFERENCE 31 Measure key (1/2) (4CH) CH1/ 2/ 3/ 4(2CH) CH1/ 2/(4CH) CH1/ 2/ 3/ 4(2CH) CH1/ 2/Vpp 1:204mV 2:24.0mVMeasureMEASUREF 1F 2F 3F 4F

Page 101

GDS-2000 Series User Manual 32 Program key (1/2) ProgramEditPlayF 1Step01.F 2ItemMenuMenu/Time/SetupF 3SaveF 5PROGRAM1 ~ 20 Select Program Edit m

Page 102 - Vertical View (Channel)

QUICK REFERENCE 33 Save/Recall key (1/9) Save/RecallDefaultSetupF 1DisplayRefs.F 2SaveSetupF 3SaveWaveformF 4MoreF 5SAVE/RECSaveImageF 1SaveAllF 2

Page 103 - Coupling

GDS-2000 Series User Manual 34 Save/Recall key (3/9) SaveSetupF 1DestinationUSBF 3SaveF 4FileUtilitiesF 5Memory/USB(USB only)To File UtilitiesSav

Page 104 - BW Limit

QUICK REFERENCE 35 Save/Recall key (5/9) SaveAllF 1SaveF 4FileUtilitiesF 5Save AllDestinationUSBF 3(USB only)To File UtilitiesInk Saver OffF 2 O

Page 105

GDS-2000 Series User Manual 36 Save/Recall key (7/9) RecallWaveformF 1RecallF 4FileUtilitiesF 5Recall WaveformDestinationF 3(USB only)To File Uti

Page 106 - Trigger

QUICK REFERENCE 37 Save/Recall key (9/9) SelectF 1DeleteF 4PreviousMenuF 5File UtilitiesRenameF 3NewFolderF 2FILE UTILSEnterCharacterF 1SaveF 4Pre

Page 107 - Trigger parameter overview

GDS-2000 Series User Manual 38 Trigger key (2/5) TypePulseF 1SourceCH1F 2ModeAutoF 3Slope /CouplingF 5TRIGGERAuto/ Normal/SingleWhen <20.0nsF

Page 108

QUICK REFERENCE 39 Trigger key (4/5) LF/ HF/ OffOn/ Off//SlopeF 1CouplingF 2Rejection OffF 3PreviousMenuF 5TRIGGERNoise Rej OffF 4Slope/Coupli

Page 109 - Setup Holdoff and Auto level

GDS-2000 Series User Manual Save ... 128Recall ...

Page 110 - Use edge trigger

GDS-2000 Series User Manual 40 Utility key (1/9) / / Off English/Chinese(S)Chinese(T)/SpanishKorean/ etcUtilityHardcopyMenuF 1InterfaceMenuF

Page 111 - By Event

QUICK REFERENCE 41 Utility key (3/9) FunctionSave AllGrayPortrait(Printer only)5 ~ 75(Printer only)Color Portrait/Gray PortraitSaveImage/SaveAll/

Page 112

GDS-2000 Series User Manual 42 Utility key (5/9) (4CH) CH1/ 2/ 3/ 4(2CH) CH1/ 2Template EditSourceCH1ViolatingStopGo-NoGo OffRatio:00STOP / STO

Page 113 - Use video trigger

QUICK REFERENCE 43 Utility key (7/9) UtilityProbeCompMenuF 1Time SetMenuF 2MoreF 5UTILITYTo Probe menuTo Time set menu Go to Probe Compensation me

Page 114 - Use pulse width trigger

GDS-2000 Series User Manual 44 Utility key (9/9) Select date/time setting F1 Select day/month/year F2→VAR Select hour/minute F2 →VAR Save d

Page 115

QUICK REFERENCE 45 Default Settings Here is the factory installed panel setting which appears when pressing the Save/Recall key→F1 (Default Setup).

Page 116

GDS-2000 Series User Manual 46 Built-in Help The Help key shows help contents. When a functional key is pressed, simple explanations of its major

Page 117 - Portuguese

MEASUREMENT 47 MEASUREMENT Basic measurement Channel activation ... 48 Auto Set ...

Page 118

GDS-2000 Series User Manual 48 Basic Measurement This section describes the basic operations required in capturing and viewing the input signal. F

Page 119 - SAVE/RECALL

MEASUREMENT 49 Auto Set Background Auto Set function automatically configures the panel settings to position the input signal to the best viewing

Page 120 - File Format/Utility

SAFETY INSTRUCTIONS 5 SAFETY INSTRUCTIONS This chapter contains important safety instructions that you must follow when operating GDS-2000 and when

Page 121 -  vertical scale

GDS-2000 Series User Manual 50 Run/Stop Background By default, the waveform on the display is constantly updated (Run mode). Freezing the wavefor

Page 122

MEASUREMENT 51 Horizontal position/scale For more detailed configuration, see page95. Set horizontal position The horizontal position knob moves t

Page 123 - Utilities

GDS-2000 Series User Manual 52 Vertical position/scale For more detailed configuration, see page102. Set vertical position To move the waveform u

Page 124

MEASUREMENT 53 Probe compensation signal Background This section introduces how to use the probe compensation signal for general usage, in case th

Page 125

GDS-2000 Series User Manual 54 3. Press F5 (More) twice. MoreF 5 MoreF 5 4. Press F1 (Wave type) repeatedly to select the wave type. 5. (Fo

Page 126 - Quick Save (HardCopy)

MEASUREMENT 55 Automatic Measurement Automatic measurement function measures and updates major items for Voltage, Time, and Delay type. Measurement

Page 127 - Hardcopy

GDS-2000 Series User Manual 56 Vavg Averaged voltage of the first cycle Vrms RMS (root mean square) voltage ROVShoot Rise overshoot voltage

Page 128

MEASUREMENT 57 FFR Time between: Source 1 first falling edge and Source 2 first rising edge FFF Time between: Source 1 first falling edge and S

Page 129 - Save panel setting

GDS-2000 Series User Manual 58 F 1F 2F 3F 4F 5 Select measurement item 3. The selection menu appears. Press F1 (Source 1) repeatedly to selec

Page 130 - Save waveform

MEASUREMENT 59 Display All mode Display All mode shows and updates all items from Voltage and Time type measurement. View measurement result 1. P

Page 131 - Destination

GDS-2000 Series User Manual 6 Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or conta

Page 132 - Save display image

GDS-2000 Series User Manual 60 Cursor Measurement Cursor line, horizontal or vertical, shows the position and value of the waveform and math opera

Page 133

MEASUREMENT 61 Parameter T1 Time position of the left cursor T2 Time position of the right cursor ∆ The time distance between the left an

Page 134 - Save All

GDS-2000 Series User Manual 62 Use vertical cursor Panel operation/ Range 1. Press the Cursor key. Cursor 2. Press F1 (Source) repeatedly to se

Page 135 - Ink Saver

MEASUREMENT 63 5. Use the Variable knob to move the cursor up or down. The F5 content changes accordingly. Example V1V2V1V2 Note: FFT Math The F

Page 136

GDS-2000 Series User Manual 64 Math Operation Overview Background Math operation runs addition, subtraction, multiplication, or FFT using the inp

Page 137 - Save/Recall

MEASUREMENT 65 Flattop FFT window Frequency resolution Not good Amplitude resolution Good Suitable for... Amplitude measurement on periodic wavef

Page 138

GDS-2000 Series User Manual 66 4. (For 4CH model only) press F2 repeatedly to select the channel pairs, 1&2 or 3&4. CH1+CH2F 2 5. The

Page 139 - Ref.A On

MEASUREMENT 67 FFT Panel operation 1. Press the Math key. MATH 2. Press F1 (Operation) repeatedly to select FFT. OperationFFTF 1 3. Press F2

Page 140 - Recall panel setting

GDS-2000 Series User Manual 68 6. To move the FFT waveform vertically, press F4 (Position) and use the Variable knob. Position0.00 DivF 4 Ran

Page 141

MEASUREMENT 69 Go-NoGo Test Overview Background Go-NoGo test checks if a waveform fits inside the user-specified maximum and minimum amplitude bou

Page 142 - Recall waveform

SAFETY INSTRUCTIONS 7 Power Supply WARNING  AC Input voltage: 100 ~ 240V AC, 48 ~ 63Hz  The power supply voltage should not fluctuate more than

Page 143

GDS-2000 Series User Manual 70 Edit: Buzzer sound Panel operation 1. Press the Utility key. Utility 2. Press F3 repeatedly to select the buzze

Page 144 - Recall image

MEASUREMENT 71 Edit: Source signal 1. Press the Utility key. Utility 2. Press F5 (More). MoreF 5 3. Press F3 (Go-NoGo Menu). Go-NoGoMenuF 3

Page 145 - Ref Image

GDS-2000 Series User Manual 72 Stop The test stops when the NoGo condition is met. The buzzer does not sound. Stop+ The test stops and the buz

Page 146 - PRINT OUT

MEASUREMENT 73 Auto Creates the upper and lower boundary together from an input signal, not from internally stored waveform. Advantage: No need to

Page 147 - 3 Configure content

GDS-2000 Series User Manual 74 Max (marked as waveform “A” in thd display) Maximum boundary: RefA, W1 ~ 20 internal memory Min (marked as wave

Page 148 - 4 Printout

MEASUREMENT 75 Auto setting 1. Make sure the source signal, on which the templates are based, appears on the display. 2. Press the Utility key.

Page 149 - REMOTE CONTROL

GDS-2000 Series User Manual 76 10. When the templates are set, press F4 (Save & Create) to save it. Save & CreateF 4 Run Go-NoGo test

Page 150 - Interface Configuration

MEASUREMENT 77 5. Press F4 (Go-NoGo). The Go-NoGo test starts running and stops according to the continue/stop condition (page71). To stop the

Page 151 - Configure RS-232C interface

GDS-2000 Series User Manual 78 Program Overview Background Program function measures input signals using cursors or automatic measurement functio

Page 152 - 1 2 3 4 5

MEASUREMENT 79 Edit program This section assumes that the panel setting is already defined and saved (step 1 and 2 in the previous page). Panel op

Page 153 - Interface

GDS-2000 Series User Manual 8 (Pollution Degree) EN 61010-1:2001 specifies the pollution degrees and their requirements as follows. GDS-2000 fall

Page 154

GDS-2000 Series User Manual 80 Setup Selects the panel setup stored in the internal memory. S01 ~ S20. For panel setup store/recall details, see

Page 155

MEASUREMENT 81 Run program This section assumes that the program editing (see previous page) is completed. Panel operation 1. Press the Program k

Page 156 - BATTERY OPERATION

GDS-2000 Series User Manual 82 4. Press F3 (From/To) to select the From: side. Use the Variable knob to select the program start step: 1 ~ 20. T

Page 157

CONFIGURATION 83 CONFIGURATION Acquisition Select acquisition mode ... 85 Select waveform memory length ...

Page 158 - MAINTENANCE

GDS-2000 Series User Manual 84 Trigger Trigger type overview ... 106 Trigger parameter overview ...

Page 159 - Probe Compensation

CONFIGURATION 85 Acquisition Acquisition process samples the analog input signals and converts them into digital format for internal processing. Se

Page 160 - Auto Set

GDS-2000 Series User Manual 86 Example Normal Peak Detect Average (2 times) Average (256 times) Peak detect effect using probe comp. wavefor

Page 161

CONFIGURATION 87 6. Press the Acquire key. Acquire 7. Press F2 (Peak Detect) or F1 (Normal) and see that in the Peak detection mode, spike nois

Page 162

GDS-2000 Series User Manual 88 2. Press F5 (Mem Leng) to select the memory length (points), short or long. Mem Leng500F 5 Range (memory point) 5

Page 163

CONFIGURATION 89 Long memory effect using probe comp. waveform 1. One of the probe compensation waveform can demonstrate long memory mode. Connect

Page 164 - APPENDIX

SAFETY INSTRUCTIONS 9 Power cord for the United Kingdom When using GDS-2000 in the United Kingdom, make sure the power cord meets the following safe

Page 165

GDS-2000 Series User Manual 90 Real time vs Equivalent time sampling mode Background GDS-2000 automatically switches between two sampling modes,

Page 166 - GDS-2000 Specifications

CONFIGURATION 91 Display Display menu defines how the waveforms and parameters appear on the main LCD display. Select waveform drawing (vector/dot)

Page 167 - APPENDIX

GDS-2000 Series User Manual 92 Accumulate waveform Background Accumulation preserves the old waveform drawings and overwrites new waveforms on to

Page 168

CONFIGURATION 93 Set display contrast Panel operation 1. Press the Display key. Display 2. Press F4 (Contrast). ContrastF 4 3a. Turn the Variab

Page 169 - Probe Specifications

GDS-2000 Series User Manual 94 Select display grid Panel operation 1. Press the Display key. Display 2. Press F5 (Grid type) repeatedly to sele

Page 170 - GDS-2202, GDS-2204

CONFIGURATION 95 Horizontal View This section describes how to set the horizontal scale, position, and waveform display mode. Move waveform positio

Page 171 - INDEX

GDS-2000 Series User Manual 96 Select horizontal scale Select horizontal scale To select the timebase (scale), turn the TIME/DIV knob; left (slow

Page 172

CONFIGURATION 97 Select waveform update mode Background The display update mode is switched automatically or manually according to timebase and tr

Page 173

GDS-2000 Series User Manual 98 Roll mode Updates and moves the waveform gradually from the right side of the display to the left. Manually select

Page 174

CONFIGURATION 99 Zoom waveform horizontally Panel operation/ range 1. Press the Horizontal Menu key. HORIMENU 2. Press F2 (Window) key. WindowF