Digital Storage Oscilloscope GDS-1000 Series PROGRAMMING MANUAL GW INSTEK PART NO. ISO-9001 CERTIFIED MANUFACTURER October 2007 edition This
COMMAND DETAILS 19 8BChannel / Math Command :CHANnel<X>:BWLimit... 206H19:CHANnel<X>:COUPl
COMMAND DETAILS 21 :channel<X>:invert <Boolean> :channel<X>:invert? :chan<X>:inv <Boolean> :chan<X>:inv? Param
COMMAND DETAILS 23 31B:CHANnel<X>:SCALe SetQuery Description Sets or returns the vertical scale. The scale depends on the probe attenuation
COMMAND DETAILS 25 33B:CURSor:Y<X>Position SetQuery Description Selects or returns the vertical (Y axis) cursor position. Same as: Cursor k
COMMAND DETAILS 27 36B:CURSor:SOURce SetQuery Description Selects or returns the cursor source channel. Same as: Cursor key →F1 (Source) Syntax &
COMMAND DETAILS 29 Parameter <NR1> Display contrast –10 ~ 10 Lowest (–10) to the Highest (+10) Example :display:contrast 0 Sets the displa
COMMAND DETAILS 31 11BMeasure command :MEASure:FALL... 224H31:MEASure:FOVShoot ...
COMMAND DETAILS 33 Description Returns the frequency value. Same as: Measure key → F1~F5 → F3 (Frequency) Syntax < Long > < Short > :
COMMAND DETAILS 35 50B:MEASure:RISe Query Description Returns the first pulse rising edge timing. Same as: Measure key → F1~F5 → F3 (RiseTime) Syn
COMMAND DETAILS 37 Example :measure:source 1 :measure:vamplitude? Selects Channel 1, and then measures the rise Voltage amplitude. 55B:MEASure:VAVe
TABLE OF CONTENTS 3 Table of Contents INTERFACE OVERVIEW ... 4 Rear Panel Overview ...
COMMAND DETAILS 39 Note Before using this command, select the measurement channel. See the example below. Example :measure:source 1 :measure:vmin?
COMMAND DETAILS 41 1 ~ 15 W1 ~ W15 <NR1> Reference waveform 1, 2 RefA, RefB Example :memory1:recall:waveform 1 Recalls a waveform from
COMMAND DETAILS 43 Example :ref1:display 1 Turns on the reference waveform A 68B:REF<X>:LOCate SetQuery Description Moves or returns the po
COMMAND DETAILS 45 13BTime (Horizontal) command :TIMebase:DELay... 253H45:TIMebase:SCALe...
COMMAND DETAILS 47 Description Sets or returns the width of the zoomed window. Same as: Horizontal menu key → F2 (Window) → Time/div knob Syntax &l
COMMAND DETAILS 49 Description Starts waiting for a trigger condition. Same as: Run key Syntax :run 78B:SINGle Set Description Selects the single
COMMAND DETAILS 51 Description Selects or returns the trigger mode. Same as: Trigger key → F5 Syntax < Long > < Short > :trigger:mode
COMMAND DETAILS 53 Example :trigger:type 2 :trigger:pulse:time 1 Selects the pulse trigger and sets the trigger time as 1sec 88B:TRIGger:REJect Se
COMMAND DETAILS 55 1 Video Example :trigger:type 0 Selects the edge trigger type 92B:TRIGger:VIDeo:FIELd SetQuery Description Selects the tri
COMMAND DETAILS 57 :trigger:video:type <NR1> :trigger:video:type? :trig:vid:typ <NR1> :trig:vid:typ? Parameter <NR1> Type <NR
INTERFACE OVERVIEW 5 Configuring the USB Interface PC side connector Type A, host USB connection GDS-1000 side connector Type B, slave Speed 1.1/2
COMMAND OVERVIEW 7 List of Command in Functional Order System *IDN... 13*L
COMMAND OVERVIEW 9 Time (Horizontal) :TIMebase:DELay... 45:TIMebase:SCALe...
COMMAND OVERVIEW 11 Command Page Command Page T :TRIGger:NREJ 166H51 :TIMebase:DELay 167H45 :TRIGger:PULSe:MODe 168H52 :TIMebase:SCALe 169H
COMMAND DETAILS 13 5BSystem command *IDN... 198H13*LRN ...
COMMAND DETAILS 15 6BAcquisition Command :ACQuire:AVERage..... 203H15:ACQuire:MODe ...
COMMAND DETAILS 17 Data size Indicates the data size. The data size is always 8008 (4000 points per channel). Time interval Indicates the time