Signal and Tracking Generator USG Series USER MANUAL REVISION 1.1 January 2014 ISO-9001 CERTIFIED MANUFACTURER
USG Series User Manual 8 of control devices: Any java-enabled PC: Windows, Mac or Linux PCs. Any android device that supports USB OTG (USB On The
GETTING STARTED 9 Appearance Front Face RF output terminal RF Output Terminal RF output port. Accepts RF outputs. Output: 0 ~ -30dBm Input im
USG Series User Manual 10 Signal Generator Display - Java Power indicatorPower SweepFunction ModeStart keyOffset and time dwell settingsFrequency
GETTING STARTED 11 Frequency Settings Sets the basic start and stop frequency parameter settings. It also displays the instantaneous (current) outpu
USG Series User Manual 12 Signal Generator Display – Android App Power indicatorFunction settingsMain panelStart keyFrequency range settingsSystem
GETTING STARTED 13 Function Settings Chooses the type of function to be performed by the USG: Continuous Sweep, Single Sweep, Single Point or Hoppin
USG Series User Manual 14 Time Dwell Settings The time dwell settings determine how long the signal will stay (dwell) at each frequency point. Ste
GETTING STARTED 15 Primary RF – Tracking Generator Function USG Power LevelPointGSP SettingsStopMarkNormalize SettingsContinuous SweepSingle Sweep
USG Series User Manual 16 First Time Use Instructions The following instructions will go over all installation instructions that are required to o
GETTING STARTED 17 3. From the device tree go to: Other devices>USB Serial Port The yellow error sign indicates that a driver has not been in
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied,
USG Series User Manual 18 6. The USG will now become available in the device tree under PORTS (COM & LPT). Disabling the Device Driver Signa
GETTING STARTED 19 Installing the USG Software from Google Play Description The USG software for controlling the USG as a signal generator can be f
USG Series User Manual 20 Tracking Generator Software Installation (Primary RF) Description Primary RF is used in conjunction with the USG as a t
GETTING STARTED 21 3. Primary RF will now be available in the Windows Start Menu.
USG Series User Manual 22 OPERATION Signal Generator Function ... 23 Setup ~ PC ...
OPERATION 23 Signal Generator Function The signal generator function can be controlled with PC using a java program (using Windows, Mac OS X or Linu
USG Series User Manual 24 3. Open USG_GUI_v1001.jar file (accessible on the User Manual CD). The USG_GUI_v1001 file doesn‟t need to be instal
OPERATION 25 4. By default, the USG app should load up when the Android device is connected to the USG. If the app does not automatically load
USG Series User Manual 26 Frequency Function Mode Description There are four different frequency modes that can be selected. Java App Display Fun
OPERATION 27 Selecting the Frequency Description Sets the Start and Stop frequency for the USG. Java App Display Start frequencyCurrent frequencyFr
Table Of Contents 1 Table of Contents SAFETY INSTRUCTIONS ... 2 GETTING STARTED ...
USG Series User Manual 28 Selecting the Frequency Step Size Description The step size settings determine the number of frequency points for the s
OPERATION 29 Time Dwell Description The Time Dwell setting determines the amount of time between each point in a sweep. Java App Display Time Dwell
USG Series User Manual 30 Frequency Offset Description The Frequency Offset setting will offset the frequency by ±0.05 MHz. Java App Display Freq
OPERATION 31 Selecting the Power Sweep Description Sets the power level for the start and stop frequencies. For the sweep functions, sets the powe
USG Series User Manual 32 Steps 1. To set the initial power level, press From. 2. To set the final power level press To. If you only want on
OPERATION 33 Tracking Generator Function (Primary RF Software) The Primary RF software can be used as a tracking generator for the GSP-730 to track
USG Series User Manual 34 Tracking Generator Setup Description The section will describe how to use the USG as a tracking generator for the GSP-7
OPERATION 35 3. Press Action(A) and select USG as TG for GSP-730. A new window will open for the tracking generator options. 4. Press Point an
USG Series User Manual 36 6. Under the GSP-730 panel set the basic spectrum analyzer settings: Ref.Level: -40 ~ 20 dBm Scale: 1~10 Start f
OPERATION 37 2. Press Single Sweep and perform a single sweep. It may take some time to complete a full sweep, depending on the fixed RBW (1MHz)
USG Series User Manual 2 SAFETY INSTRUCTIONS This chapter contains important safety instructions that you must follow during operation and storage
USG Series User Manual 38 Note Before performing normalization, make sure the USG has been on for at least 30 minutes to eliminate drift from eff
OPERATION 39 Save Results - CSV Description The spectrum results on GSP-730 can be saved as a CSV file. CSV File Format Center frequencyAmplitudePo
USG Series User Manual 40 Save Results - Print Description The spectrum results on the GSP-730 can be saved as a PDF or printed. Print Format Re
FAQ 41 FAQ • The USG will not connect to the PC. • Primary RF will not allow me to connect to the GSP-730. • The performance does not match the spe
USG Series User Manual 42 APPENDIX USG Specifications The specifications apply when the USG is powered on for at least 30 minutes to warm-up to a
APPENDIX 43 3rd Harmonics 0 dB Attenuation ≤ -5 dBc, typical 34.5 MHz to 2.0 GHz, fundamental ≤ -20 dBc, typical 2.0 GHz to 3.0 GHz, fundamental
USG Series User Manual 44 USG-0818 Frequency Range 800 MHz to 1.8 GHz Output Power -30 dBm to 0 dBm in 1 dB steps Internal Reference 25 MHz aging
APPENDIX 45 Amplitude Absolute Accuracy 0 dBm ± 1 dB typical at 2500 MHz, 0 dBm Output Output Flatness ± 1 dB, ref. to 2500MHz at 0 dBm Output Phase
USG Series User Manual 46 Spurious related to Resolution settings (Single Point Mode) ≤ -30 dBc, typical Resolution < 1MHz ≤ -65 dBc, typical R
APPENDIX 47 USG Dimensions 20.8102.429.5102.420.827.8Scale :mm
SAFETY INSTRUCTIONS 3 Safety Guidelines General Guideline CAUTION Do not place any heavy object on the instrument. Avoid severe impact or ro
USG Series User Manual 48 Declaration of Conformity We GOOD WILL INSTRUMENT CO., LTD. No. 7-1, Jhongsing Rd, Tucheng Dist., New Taipei City 236,
INDEX 49 INDEX Accessories ... 8 Caution symbol ... 2 Cleaning the instrument ...
USG Series User Manual 4 Storage environment Location: Indoor Temperature: -20°C to 60°C; 65°C @ 70% RH Disposal Do not dispose this instru
GETTING STARTED 5 GETTING STARTED This chapter provides a brief overview of the USB Signal Generator (hereafter referred to as „USG‟), the package c
USG Series User Manual 6 Tracking Generator Software Installation (Primary RF) .... 20
GETTING STARTED 7 USG Series Introduction The USG series signal and tracking generators can be operated as standalone continuous wave generators, or